User guide

6 Devices
IC Net 2.3
164
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
Third column Parameter for program instruction (see be-
low).
Enabled Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program page:
Instruction Parameter entry Meaning
Suppressor on, off
Switch the Dual Suppressor
on/off.
Links
The
Links tab of the 828 IC Dual Suppressor is used for Event line
selection.