User guide
6.16 793 Sample Prep Module
IC Net 2.3
155
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
Third column Parameter for program instruction (see be-
low).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program page:
Instruction Parameter entry Meaning
Program RESET, END
Set program flag. The END flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration. Additional
steps after this flag are not al-
lowed. The
RESET flag is used to
reset the parameters to the sys-
tem startup values.
Pump on, off Switch on or off the pump drive.
Step Rotate sample preparation mod-
ule to the next position.
Links
The
Links tab of the 793 Sample Prep Module is used for Event line
selection.