User guide
6.15 754 Dialysis Unit
IC Net 2.3
151
Manual
The Manual tab of the 754 Dialysis Unit window for manual opera-
tion is only available for a connected system.
Pump
<On> Start pump drive.
<Off> Stop pump drive.
Start pump with startup hardware
Automatic start of pump drive with Startup
hardware
.
Time program
On the Program tab of the 754 Dialysis Unit window a user-defined
time program can be entered. This program is started automati-
cally as defined in the Start mode window (see section 4.4.3) either
at the moment the determination is started (Start with determina-
tion
) or at the moment the sample is injected (Start with inject).
First column Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).