User guide
6.9 820 IC Separation Center
IC Net 2.3
123
Thermostat
The Thermostat tab of the 820 IC Separation Center or 820.02X0 IC
Separation Center
window contains the temperature control of the
column heater.
Only available for a 2.820.02X0 Separation Center that is con-
nected either to the 819 IC Detector and installed as 820.02X0
to IC Detector, or to the 830 IC Interface and installed as
820.02X0 to Interface.
Actual temperature Display of actual temperature within the
column heater in °C.
Set point Operating temperature of the thermostat
within the column heater in °C.
Entry Range: 10.0 … 80.0.
<On> Only active if the thermostat is off. Switch
the thermostat of the column heater on and
apply the operating temperature specified
under Set point, the button turns to <Set>.
<Set> Only active if the thermostat is on and a
new operating temperature is entered at Set
point
. Sets the new operating temperature.
<Off> Only active if the thermostat is on. Switch
the thermostat of the column heater off.