User guide
6.9 820 IC Separation Center
IC Net 2.3
121
Valve A
fill/inject Display of current valve position.*
<FILL> Switch valve A to "fill" position.
<INJECT> Switch valve A to "inject" position.
Valve B
fill/inject Display of current valve position.*
<FILL> Switch valve B to "fill" position.
<INJECT> Switch valve B to "inject" position.
Suppressor
in position Display of current suppressor status.*
<STEP> Switch suppressor module to next position.
Auto Step *
fill, inject, off Display of current auto step status.
Entry field Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection: fill, inject, off
* only available for 820 IC Separation Center connected to 819
Time program
The Program tab is only available for a 820.0XX0 Separation Center
connected to a 830 IC Interface. The 820 IC Separation Center
that is connected to a 819 IC Detector is controlled by the time
program of the detector.
On the Program tab of the 820.0XX0 Separation Center window a
user-defined time program can be entered. This program is
started automatically as defined in the
Start mode window (see
section 4.4.3) either at the moment the determination is started
(
Start with determination) or at the moment the sample is injected
(
Start with inject).