Instruction Manual
6 Devices
IC Net 2.2
78
Oven Display of current oven state.
Range Display of current measuring range.
Full scale Display of current full scale range.
Offset Display of current offset.
Cycle potentials Graphical display of cycle potentials.
Method Display of working potentials set in the
Method data frame.
Actual Display of current working potentials.
<Set> Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (*.smt) as long as the file is
not saved.
Scan mode
The Scan mode subtab of the Method parameters tab on the 817
Bioscan
window is used for parameter setting and curve display
for the scan mode. This window is only available if Mode has been
set to Scan on the Manual tab.
Parameters
Method data
Cell potential Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.
E1 Start potential for sweep.
Range: -2 ... +2 V.
E2 End potential for sweep.
Range: -2 ... +2 V.
Scan cycle Definition of scan cycle:
full double scan E1 ... E2 ... E1
half single scan E1 ... E2
Selection: full, half
Scan rate Definition of scan rate in mV/s.
Selection: 1, 2, 5, 10, 20, 50 mV/s
Oven Switching on/off column oven and setting
operating temperature.
Range: 15 ... 60 °C
Range Selection of measuring range:
pA 10 ... 5000 pA.
nA 0.1 ... 50 nA.
µA 0.01 ... 5 µA.