Instruction Manual
6.5 817 Bioscan
IC Net 2.2
75
Valve Manual switching of the injection valve at
the 812 Valve Unit:
<Fill> Switch injection valve to "Fill" position (con-
trol line Cont1 is set to 0).
<Inject> Switch injection valve to "Inject" position
(control line Cont1 is set to 1).
Analog output
Manual setting of the analog output signal
to zero.
<Zero> Trigger the autozero function.
Actual
Display of actual values.
Mode Display of current detector mode.
Cell Display of current cell state.
Oven
Display of current ovent state.
Control
Display of current control line settings.
DC mode
The DC mode subtab of the Method parameters tab on the 817 Bio-
scan
window is used to set the parameters for the DC mode. This
window is only available if Mode has been set to DC on the Manual
tab.
Method data
Cell potential Electric potential applied to the amperomet-
ric detector cell in V
Range: -2 ... +2 V.