Instruction Manual
6 Devices
IC Net 2.2
66
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Valve Inject, Fill
Switch injection valve to "INJECT"
or "FILL" position.
FullScale 50, 250, 1000
ยตS/cm
Set full scale range to the se-
lected value.
ICPump on, off Switch on or off the high-pressure
pump.
Flow 0.2 ... 2.5 mL/min Set flow rate of the high-pressure
pump to the desired value.
Pmax 0.0 ... 25.0 MPa Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin 0.0 ... 25.0 MPa Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote 0, 1, *, p Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter 1, 0, p or *. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter 1, 0, p or *.
Program END, RESET The END flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The
RESET flag is used to reset the
parameters to the system startup
values.
Suppressor Switch suppressor module to the
next position.
Peristaltic on, off Switch on or off the peristaltic
pump.