Instruction Manual

4 Systems
IC Net 2.2
30
Ident User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the
chromatogram window and in the Chroma-
togram
open window.
Calibration level Calibration level (0 = sample; 1n = cali-
bration solutions).
Info 1 / Info 2 Sample description.
Volume Injected volume in µL.
Dilution Dilution of the sample.
Vial number Autosampler vial position to take sample
from.
Amount Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
Amount
s
/ Amount
c
Internal standard amount
Concentration of the internal standard
component for relative concentration calcu-
lations.
Date/time when... Date and time of sample collection (the de-
fault values are equal to the date and time
when the chromatogram starts).
The Verify sample option is disabled if determinations are made
with an active sample queue.