Instruction Manual
4.3 System functions
IC Net 2.2
29
4.3.3 Start/stop determinations
SYSTEM / Control / Start determination
Start determination using the settings of the selected system.
At this start command, the System startup values are set at the de-
vices. The time programs of the different devices and the data re-
cording are started either immediately (Start with determination) or
after switching the injection valve to the "Inject" position (Start with
inject
) as set in the Start mode window.
SYSTEM / Control / Stop determination
Stop running determination. Data acquisition and time pro-
grams are terminated immediately. The recorded chromatogram
is saved automatically if the Save chromatogram after the run option
on the
Passport / Processing tab is enabled.
Alternatively the determination can be stopped by clicking the
icon of the chromatogram window. In this case the user is asked
always if the determination should be saved or not.
If the Auto restart option is enabled or if a sample
queue is active, a new determination is started immedi-
ately after the running determination is stopped. If you
want to avoid this, disable the Auto restart option or
Pause the sample queue.
SYSTEM / Control / Stop data acquisition
Stop data acquisition of the running determination immediately
and save the recorded chromatogram automatically if the Save
chromatogram after the run
option on the Passport / Processing tab
is enabled. The time programs of the running determination are
continued normally.
4.3.4 Options for determinations
SYSTEM / Control / Auto restart
If this option is enabled, a new determination is started automati-
cally using the current system after the preceding determination
has been finished normally or stopped manually.
The Auto restart option is disabled if determinations are made
with an active sample queue. This option allows to make an infi-
nite batch cycle with the current system
SYSTEM / Control / Verify sample
If this option is enabled, the Edit sample description window is
opened automatically at the start of each determination for entry
of the following sample information: