User Manual
6.5 817 Bioscan
IC Net 2.1
73
Method data
Cell potential
Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.
E1
Start potential for sweep.
Range:
-2 ... +2 V
.
E2
End potential for sweep.
Range:
-2 ... +2 V
.
Scan cycle
Definition of scan cycle:
full
double scan
E1 ... E2 ... E1
half
single scan
E1 ... E2
Selection:
full, half
Scan rate
Definition of scan rate in mV/s.
Selection:
1, 2, 5, 10, 20, 50 mV/s
Oven
Switching on/off column oven and setting
operating temperature.
Range:
15 ... 60 °C
Range
Selection of measuring range:
pA 10 ... 5000 pA
.
nA 0.1 ... 50 nA
.
µA 0.01 ... 5 µA
.
Full
scale
The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring
Range
.
Offset
Offset of the zero point for the analog out-
put signal.
Selection:
-50 ... +50 %
(in steps of 10 %)
Actual
E1, E2
Display of current start and end potentials
in V.
Scan cycle
Display of current scan cycle mode.
Scan rate
Display of current scan rate in mV/s.
Oven
Display of current oven state.
Range
Display of current measuring range.
Full
scale
Display of current full scale range.
Offset
Display of current offset.
<Set>
Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (
*.smt
) as long as the file is
not saved.