Owner's manual
6 Devices
IC Net 2.0
86
Manual injection
Sample vial number
Number of sample vial to be used for man-
ual injection.
<Select> Selection of the vial number (1...128).
Count of injections per vial
Number of injections per vial (1...4).
<Select> Selection of the injection number (1...4).
Needle lowering type
Injection number Position where the needle is moved to in
case of multiple injections (1...4).
Absolute position Position where the needle is moved to in
case of single injections (1...4).
Rinse type
Standard shift Use previous or next vial for rinsing.
Nonstandard shift For every sample, skip this number of vials
and use this vial for rinsing.
No rinse between different sample vials
No rinsing.
Data acquisition started by software
If this option is enabled, the data acquisi-
tion is started with "Inject". No cable is re-
quired between the injection valve and the
interface.
If this option is disabled, the 6.2115.070
cable is needed to start data acquisition.
<Inject> Start injection.
For detailed information about the 750 Autosampler see 750 In-
structions for Use.
Links
The Links tab of the 750 Autosampler is used for COM port selec-
tion and settings (details see section 5.2.4).