Owner's manual
6 Devices
IC Net 2.0
72
Program
The Program tab of the SDU window contains a time program for
the pump control system to define eluent gradients.
Flow Overall flow of pump control system in
mL/min. This value is also available for the
WATCH WINDOW. This field is only available
if the Show flow column option is disabled.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
Stop pumps after the run
Stop pumps after the determination.
Program table
Number Row number in the program table (read-
only).
Time Time at which the flow and/or gradient is
changed.
Entry range: 0.0 ... 99999 min
Flow Overall flow of pump control system in
mL/min.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
%B Individual flow rate for pump B in percent of
total flow.
Entry range: 0 ... 100 %
%C Individual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %