Owner's manual

6.6 Metrohm solvent delivery unit
IC Net 2.0
71
Pressure Live display of current pressure in MPa.
This field is also available for the WATCH
WINDOW.
Flow Live display of current flow in mL/min. This
field is also available for the WATCH WIN-
DOW.
New flow A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
Min/Max pressure The minimum and maximum pressure limit
for the individual pump units can be set in
steps of 0.1 MPa.
Entry range: 0.0 ... 50.0 MPa.
%B Individual flow rate for pump B in percent of
total flow.
Entry range: 0 ... 100 %
%C Individual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %
%D Individual flow rate for pump D in percent of
total flow.
Entry range: 0 ... 100 %
Stop after Stop time for all pump drives.
Entry range: 0 ... 99999 min
<Set> Send current parameters immediately to
the pump control system. Parameters are
not stored in the system file (*.smt) as long
as the file is not saved.
<Start> Start all pump drives.
<Stop> Stop all pump drives.