Owner's manual
6 Devices
IC Net 2.0
66
Suppressor
in position Display of current suppressor status.
<STEP> Switch suppressor module to next position.
Auto Step
fill, inject, off Display of current auto step status.
Entry field Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection: fill, inject, off
Links
The Links tab of the 733 IC Separation Center is used for COM port
selection and settings (details see section 5.2.4).
6.5 709 IC Pump
6.5.1 709 IC Pump features
The Metrohm 709 IC Pump is a serial dual piston high pressure
pump for ion chromatography. The two following versions are
available:
• 2.709.0010 IC Pump (standard version)
• 2.709.0110 IC Pump (metal-free version)
The 709 IC Pump, which is normally operated using the keypad
of the instrument or via 732 IC Detector, can be fully operated
using the «IC Net» program. It can be connected either to the 732
IC Detector using the 6.2125.060 cable, to the 762 IC Interface
using the 6.2134.080 or 6.2134.090 cable, or to a PC COM port
using the 6.2134.040 or 6.2134.100 cable. If it is connected to the
732 IC Detector, the Isocratic pump 709 for 732 device must be se-
lected for the installation of the 709 IC Pump in a system. If it is
connected to the 762 IC Interface, the Isocratic pump 709 device
must be selected for the installation of the 709 IC Pump in a sys-
tem.
Examples for interconnections with the 709 IC Pump can be
found in the 762 Instructions for Use.
The cable connection between 709 IC Pump and 732 IC De-
tector, 762 IC Interface or PC must be made before switching
on the instrument. In order to operate the 709 IC Pump with «IC
Net», the external control must be switched on using the [ Ext. ]
key.
For detailed information about the 709 IC Pump see 709 Instruc-
tions for Use.