Owner's manual

6.4 733 IC Separation Center
IC Net 2.0
65
If the system is connected and the 733 icon is clicked with the
right mouse button, the following menu appears:
Open Open the 733 IC Separation Center window for parame-
ter settings (this window can also be opened by dou-
ble-clicking the icon).
Unlink Delete the 733 icon from the system.
6.4.3 733 IC Separation Center window
733 icon / Open
The 733 IC Separation Center window for parameter settings is
opened by selecting this menu option with the right mouse button
or by double-clicking the 733 icon in the SYSTEM window. It con-
sists of the two tabs IC733 and Links.
IC733
The IC733 tab of the 733 IC Separation Center window is only avail-
able for a connected system. The parameters and buttons for
manual operation depend on the type of instrument connected.
Configuration Display of current instrument configuration.
valve A 733 IC Separation Center with 1 injector.
valve A + valve B 733 IC Separation Center with 2 injectors.
valve A + suppressor
733 IC Separation Center with 1 injector
and 1 suppressor module.
Valve A
fill/inject Display of current valve position.
<FILL> Switch valve A to "fill" position.
<INJECT> Switch valve A to "inject" position.
Valve B
fill/inject Display of current valve position.
<FILL> Switch valve B to "fill" position.
<INJECT> Switch valve B to "inject" position.