Owner's manual

6.3 732 IC Detector
IC Net 2.0
61
The Program tab contains the two following subpages:
Program Main time program with all program
steps.
Remote configuration Possibility for creation of user-defined
remote commands.
Program
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Remote configura-
tion tab.
Third column Parameter for program instruction (see
below).
ENABLED Enable program start (a disabled program
is not started).
Will be executed by Selection if the time program will be exe-
cuted by the PC or the 732 unit.
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).