Owner's manual

6 Devices
IC Net 2.0
58
6.3.3 732 IC Detector window
732 icon / Open
The 732 IC Detector window for parameter settings is opened by
selecting this menu option with the right mouse button or by
double-clicking the 732 icon in the SYSTEM window. It consists of
the four tabs Display, Method parameter, Program, and Links.
Display
The Display tab of the 732 IC Detector window is only available for
a connected system. It displays the current measurement val-
ues and parameters settings of the 732 IC Detector.
Conductivity Live display of measured conductivity
(auto-zero value) in µS/cm. This field is
available for the WATCH WINDOW. The
watch window color is changed to the Out
of range color if the measured value is more
than 80% outside the Full Scale range.
abs. Cond. Live display of absolute conductivity value
in µS/cm. This field is available for the
WATCH WINDOW. The watch window color is
changed to the Out of range color if the
measured value is bigger or equal to 110%
of Range.
Ion Display of ion settings:
Anion: polarity = +, temperature coeff. = 2.5
Cation: polarity = –, temperature coeff. = 1.5