Owner's manual

5 Interfaces
IC Net 2.0
50
Hardware error Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector Leak detector has detected solvent in the
instruments interior.
Remote line Set remote output lines 18.
Selection: 0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse output)
* (leave line in current status)
PowerOn values Display of the power on startup values for
remote output lines set on the Hardware tab.
Pulse length Length of a pulse in ms.
Alarm stops
The Alarm stops tab of the hardware settings window defines the
events for which the 761 Compact IC is stopped immediately. At
an alarm stop, high-pressure pump and peristaltic pump are
stopped immediately, the running determination and the active
sample queue are also stopped.
Event Events for alarm stop:
Pump stop Pump stopped because pressure limits are
exceeded.
Hardware error Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Leak Leak detector has detected solvent in the
instruments interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without con-
nection to the PC.
Power on Temporary power failure at the 761 Com-
pact IC hardware.