5.4.5 Hardware settings
Selection of the Hardware menu item of the 761 icon using the right
mouse button opens the hardware settings window consisting of
the three tabs Hardware, Outputs and Alarm stops.
Hardware
The Hardware tab of the hardware settings window defines gen-
eral parameters which are set automatically at power on of the
761 Compact IC.
Remote lines after power on
The remote output lines 18 are set to this
values after power on or a manual stop with
Shutdown hardware.
Selection: 0, 1
Thermostat Operating temperature of the conductivity
cell.
Selection: 25, 30, 35, 40, 45 °C, off
Thermostatting functions only if the ambient tem-
perature is at least 5 °C lower than the operating tem-
perature. It normally takes 3060 min after power on
until a temperature stability of ± 0.01 °C is attained.
Flow correction Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range: 0.9 ... 1.09
The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:
Flow correction =
Displayed flow rate
Measured flow rate