Owner's manual
5.4 761 Compact IC
IC Net 2.0
45
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Remote configura-
tion tab.
Third column Parameter for program instruction (see
below).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Valve Inject, Fill Switch injection valve to "INJECT"
or "FILL" position.
FullScale 50, 250, 1000
µµS/cm
Set full scale range to the se-
lected value.
ICPump on, off Switch on or off the high-pressure
pump.
Flow 0.2 ... 2.5 mL/min Set flow rate of the high-pressure
pump to the desired value.
Pmax 0.0 ... 25.0 MPa Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin 0.0 ... 25.0 MPa Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote 0, 1, *, p Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter 1, 0, p or *. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter 1, 0, p or *.