Owner's manual

5 Interfaces
IC Net 2.0
44
Pressure max, MPa Startup value for maximum pressure limit
for high-pressure pump.
Entry range: 0.0 ... 25.0 MPa
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump.
Entry range: 0.0 ... 25.0 MPa
Full scale, µµS/cm Startup value for full scale range.
Selection: 50, 250, 1000 µµS/cm
Remote lines Startup value for remote line settings 18.
Selection: 0, 1
Time program
On the Program tab of the 761 Compact IC window a user-defined
time program for the remote output lines of the 761 Compact IC
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (Start with determination) or at the
moment the sample is injected (Start with inject).
The Program tab contains the two following subpages:
Program Main time program with all program
steps.
Remote configuration Possibility for creation of user-defined
remote commands.
Program
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.