Owner's manual
5.4 761 Compact IC
IC Net 2.0
43
IC pump High-pressure pump.
<On> Start high-pressure pump.
<Off> Stop high-pressure pump.
Peristaltic pump Peristaltic pump.
<On> Start peristaltic pump.
<Off> Stop peristaltic pump.
System startup values The system startup values are sent and
applied to the 761 Compact IC each time
the system is connected, a determination is
started, or the values are sent manually
with <Send to unit>.
Flow, mL/min Startup value for flow rate of the high-
pressure pump.
Entry range: 0.20 ... 2.50 mL/min
Pressure max, MPa Startup value for maximum pressure limit
for high-pressure pump. This limit is con-
trolled even without connection to the PC.
Entry range: 0.0 ... 25.0 MPa
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range: 0.0 ... 25.0 MPa
Full scale, µµS/cm Startup value for full scale range.
Selection: 50, 250, 1000 µµS/cm
Remote lines Startup value for remote line settings 1…8.
Selection: 0, 1
Initial
The Initial tab of the 761 Compact IC window for settings of system
startup values is only available for a disconnected system.
Flow, mL/min Startup value for flow rate of the high-
pressure pump.
Entry range: 0.20 ... 2.50 mL/min