Owner's manual

5.3 771 Compact Interface
IC Net 2.0
39
Range Range for AD converter.
Example: converts ± 2500 mV to 2
24
bits
Entry range: 2500, 1250, 625, 312.5, 156.25,
78.125, 39.062 mV
<ASK> Read current parameters from 771 Com-
pact Interface.
<SET> Send current parameters to 771 Compact
Interface.
RS settings
The RS settings tab of the 771 Compact Interface window contains
RS232 parameters for the 771 Compact Interface.
Baudrate Baud rate of the device.
Selection: 1200...115200
Default value: 57600
Parity Parity check.
Selection: None, Even, Odd
Default value: None