Owner's manual
5.2 762 IC Interface
IC Net 2.0
35
Time program
On the Program tab of the 762 System [#] window a user-defined
time program for the event output lines of the 762 IC Interface
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (Start with determination) or at the
moment the sample is injected (Start with inject).
The Program tab contains the two following subpages:
Program Main time program with all program steps.
Events configuration Possibility for creation of user-defined re-
mote commands.
Program
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Events configura-
tion tab.
Third column Parameter for program instruction (see
below).
ENABLED Enable program start (a disabled program
is not started).