Owner's manual

5.2 762 IC Interface
IC Net 2.0
31
5.2 762 IC Interface
5.2.1 762 IC Interface features
The Metrohm 762 IC Interface is available as a one system ver-
sion (2.762.0010) or a two system version (2.762.0020). Each
system can control a complete ion chromatography or HPLC
equipment and contains the following parts:
Two 24-bit Analog-to-Digital Converters (often abbreviated
as ADC) to acquire two analog signals from two different de-
tectors simultaneously (for example conductivity detection and
UV detection).
One external start input.
One start button at the interface front panel.
RS232 communication ports to control up to eight devices
of a complete ion chromatography equipment.
Seven Event output lines.
The 762 IC Interface is normally connected to a PC COM port
using the 6.2134.100 connecting cable.
For detailed information about the 762 IC Interface see 762 In-
structions for Use.
5.2.2 762 icon on the toolbar
The 762 icon is available on the toolbar if a 762 IC Interface has
been installed with the new system wizard or by using the <Install
device> option of the WORKPLACE window (see section 5.1.2). By
clicking this icon the 762 IC Interface window for parameter set-
tings opens.
5.2.3 762 icons in the system window
The 762 System 1 and 762 System 2 icons are available in the SYS-
TEM window if the 762 System [1] or 762 System [2] have been in-
stalled with the Install existing devices option (see section 6.1.3).
By clicking this icon using the right mouse button, the following
menu items can be selected:
Open Open the 762 System [#] window for programming the
event output lines (this window can also be opened by
double-clicking the icon).
Unlink Delete the 762 icon from the system.