Owner's manual
7.6 Report output
IC Net 2.0
139
Concentration of std Concentration of the internal standard
component for relative concentration cal-
culations. This value is stored in This run
column of the concentration table.
Total % for normalization
A value to which the sum of concentrations
is normalized. It is used for Response nor-
malization and Normalized concentrations
quantification methods. The default value is
100.
Printing order Determines the order of the components in
the peak table.
By peaks Lists run results for all peaks detected.
Unidentified peaks (corresponding to the
universal component) are included, but
missing components are not listed.
By components Missing components are always included
into the report, even those with concentra-
tion equal to zero. In the case of presence
of universal component the summary for all
unidentified peaks is presented as a single
line.
<Customize> Calls up a box with a list of columns that
can be included into the report (see section
7.6.8). It is available only if Quantification
method = Custom has been selected.
Report all peaks Checkbox that includes all peaks irrespec-
tive of the concentration of the substance.
Otherwise the report generator excludes
lines with zero concentration of the compo-
nent. It is available only if Quantification
method = Custom has been selected.
Groups If this checkbox is checked, a separate
peak table report is put out for every group
specified in the component table.