Owner's manual
7 Methods
IC Net 2.0
118
Force horizontal Cancel horizontal
base back base back
Force horizontal base back
Set horizontal baseline for a single peak.
The baseline is drawn backwards from the
peak end point. The intersection of the
baseline with the signal is defined as peak
start point.
Cancel horizontal base back
Clear Force horizontal base back mode.
Enable baseline Disable baseline
penetration penetration
Enable baseline penetration
Allow crossing of the signal by the base-
line.
Disable baseline penetration
Clear Enable baseline penetration mode.
Integration events for integration parameters
Set width Sets new Width parameter that supersedes
default linear growth of expected peak.
Set slope Sets new Slope value.
Set min height Sets new Minimum peak height value.
Set rider ratio Sets new value of Rider ratio parameter.