Owner's manual
7.4 Integration
IC Net 2.0
115
If an integration event is added or modified, the Edit integration
event window is opened where the following parameters can be
set:
Time Start time for integration event in min.
Event List box with integration events to be se-
lected (see below).
Value Parameter value for selected integration
event. This field appears for events that
demand input of an additional parameter.
Integration events for peak detection
Disable detection Set this mode to stop detection of new
peaks. If a peak is available at the moment
of the event it is either finished (downslope
peaks) or rejected (upslope peaks).
Enable detection Clear Disable detection mode.
Enable negative peaks Enable detection of negative peaks. In
some cases this mode may result in insta-
bility of detection algorithm.
Disable negative peaks Disable detection of negative peaks. This
event does not influence negative peaks
which already started.
Disable peak reject Set this mode, if a small peak should not
be rejected because of its flat apex.
Enable peak reject Clear Disable peak reject mode.