114
<Suggest> This function sets up reasonable integration
parameters in the following way:
Slope is set to a reasonable value.
Asymmetry is set to a reasonable value.
Width and Broadening are calculated by
fitting a straight line through the meas-
ured peak halfwidth values vs. retention
time. Width is determined as the y value
of this line at the Delay time, Broadening
is calculated as the y value at the end of
the chromatogram divided by the y value
at the Delay time. If this procedure does
not provide adequate result,
Width is set equal to the minimum value
of peak width for the chromatogram,
Broadening is set to a reasonable value.
7.4.3 Events
Tab Events of Integration parameters window for definition of inte-
gration events. Integration events have a higher priority than inte-
gration parameters and are used for fine tuning of the integration
process. They should be used only in the case when problems
cannot be solved by tuning a set of parameters from the integra-
tion parameters window.
Number of events Number of integration events (read-only).
<Add> Add event to the list.
<Modify> Edit the selected event.
<Delete> Delete the selected event.
<Apply> Accept integration parameters and perform
reintegration.