Owner's manual
6 Devices
IC Net 2.0
92
Ctrl "control
command"
Set the specified remote output
signal. In addition to the normal
control commands (details see 766
Instruction for Use) special
commands defined on the Configu-
ration tab can be used.
Wait 1...9999 s Waiting time.
StartDA Start data acquisition.
Configuration
The Configuration tab of the 766 IC Sample Processor window con-
tains parameters settings on the five subtabs Rack, Beaker posi-
tions, Needle positions, Control, and Scan.
Rack
Rack definitions
Code Binary ID code for rack identification (read-
only).
Type Predefined rack type (read-only).
Settings
Maximum lift way Definition of lowest lift position possible in
mm.
Entry range: 0...125 mm
<Apply> Send parameters to 766 IC Sample Proc-
essor.
<Ask> Read current parameters from 766 IC
Sample Processor.