User guide
4.5 10-port valve
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
920 Absorber Module
Connect
capillary/tub-
ing at port
…
… to
10 Waste container
4.5.2 Connecting the bottle with the standard solution
The bottle with the standard solution is connected to the 10-port valve.
Accessories For this step you need:
■ Clear glass bottle 100 mL (6.1608.050) filled with standard solution
■ Bottle attachment (6.1602.150), contains 3 pressure screws
Connecting the bottle with the standard solution
6.1608.050
6.1602.150
6.1602.150
6.1803.160
1
Connecting the aspirating capillary for the standard solution
■ Slide a pressure screw over the loose end of the capillary connec-
ted to port 2 of the 10-port valve (13-3).
■ Thread the capillary through one of the three openings on the
bottle attachment.
■ Screw the bottle attachment onto the bottle.
■ Push the capillary far enough into the bottle that it reaches the
bottom of the bottle. Then fix it in place with a pressure screw.