User guide

■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
57
3
Connecting port 2
Cut off a piece from the PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160) that is approx. 80 cm long using the capillary cutter.
Screw this piece of capillary in place at port 2 using a long pres-
sure screw.
4
Connecting port 3
Cut off a piece from the PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160) that is approx. 80 cm long using the capillary cutter.
Screw this piece of capillary in place at port 3 using a long pres-
sure screw.
5
Sealing ports 4 through 7
Ports 4 through 7 are not used.
Seal each of ports 4, 5, 6 and 7 with a threaded stopper.
6
Connecting port 8
Cut off a piece from the PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160) that is approx. 60 cm long using the capillary cutter.
Screw this piece of capillary in place at port 8 using a long pres-
sure screw.
7
Connecting port 9
Cut off a piece from the PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160) that is approx. 5 cm long using the capillary cutter.
Screw this piece of capillary in place at port 9 using a long pres-
sure screw.
8
Connecting port 10
Screw an entire PTFE capillary, 0.75 mm ID / 3 m (6.1803.160) in
place at port 10 using a long pressure screw.
Next step
Connect
capillary/tub-
ing at port
… to
1 6-port valve (see Chapter 4.6, page 61)
2 Bottle with standard solution (see Chapter 4.5.2, page 58)
3 Bottle with check standard (see Chapter 4.5.3, page 59)
8 Absorber tube (see Chapter 4.7, page 67)