User guide
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
55
4
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 80 cm. Connection between the
10-port valve and the bottle with the check
standard.
5
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 60 cm. Connection between the
10-port valve and the absorber tube.
6
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 5 cm. For aspirating air segments.
7
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Connection between the 10-port valve and
the waste container.
4.5.1 Connecting the ports
The transfer tubing and capillaries for all of the liquids being used are con-
nected to the 10-port valve.
Connectors
The ports of the 10-port valve are used as follows:
Port … … is used for … … is connected to
Central Aspirating and discharging differ-
ent solutions
Transfer tubing
1 Outlet for sample, standard solu-
tion, check standard
6-port valve
2 Inlet for standard solution Bottle with standard solution
3 Inlet for check standard Bottle with check standard
4-7 Not used –
8 Inlet for sample Absorber tube
9 Aspirating air segments –
10 Outlet for waste Waste container
Accessories The following accessories have to be connected to the 10-port valve:
Port Install this accessory
Central ■ Adapter UNF 10/32 outer / M6 inner (6.2744.200)
■ Transfer tubing (6.1562.130)
1
■ PEEK capillary, 0.5 mm ID / 15 cm (6.1831.040) (13-2)
■ Pressure screw (6.2744.090)
2 ■ PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 80 cm
(13-3)
■ Pressure screw (6.2744.090)
3 ■ PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 80 cm
(13-4)
■ Pressure screw (6.2744.090)
4-7 Threaded stopper (6.2744.220)