Owner's manual
3.3 Mounting the extension module below the IC instrument
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
872 Extension Module IC Pump
9
Mounting the drainage tubing
Reconnect the drainage tubing to the drainage tubing connector of
the bottle holder.
Possibly, a longer section of silicone tubing 6.186.020 must be cut to
fit and mounted (see also the manual for the IC instrument).
3.3 Mounting the extension module below the IC instru-
ment
1
Switching off the IC instrument
Switch the IC instrument off and disconnect the mains cable.
2
Clearing the bottle holder
If there are bottles and other things on the bottle holder, remove
them.
3
Disconnecting all connections on the rear of the instrument
■ Disconnect the mains cable,
■ Disconnect the MSB cable,
■ Disconnect the USB cable,
■ Disconnect the leak sensor,
■ Remove the drainage tubings.
4
Removing the detector(s)
Disconnect the detector cable(s) and remove the detector(s) from the
IC instrument (see manual for the IC instrument).
5
Removing the base tray
■ Tilt the IC instrument sideways and lay it down flat.
■ Loosen the cylinder screws with a 6.2621.100 3 mm hexagon
key.
■ Remove base tray.