865 Dosimat plus Manual 8.865.
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 865 Dosimat plus Manual 8.865.8005EN 09.
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above. Documentation in additional languages can be found on http://products.metrohm.com under Literature/Technical documentation.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1 1.1 Instrument description ......................................................... 1 1.1.1 Dosing modes ......................................................................... 1 1.1.2 Connectors .............................................................................. 1 1.2 Intended use ......................................................................... 2 1.3 About the documentation .............................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 4.2.4 4.2.5 Entering text and numbers ..................................................... 19 Selecting from a selection list ................................................. 20 4.3 Methods .............................................................................. 20 4.3.1 Method templates ................................................................. 20 4.3.2 Creating a new method ......................................................... 21 4.3.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 6.4 Pipetting and diluting (LQT) .............................................. 54 6.4.1 General description ................................................................ 54 6.4.2 Selecting a solution ............................................................... 56 6.4.3 Definitions of the sample transfer .......................................... 56 6.4.4 Dosing parameters ................................................................ 57 6.4.5 Reports .....
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 10.2 Quality Management Principles ........................................ 78 10.3 Warranty (guarantee) ......................................................... 79 11 Accessories Index VI ■■■■■■■■ 80 11.1 Scope of delivery 865 Dosimat plus 2.865.0010 ............. 80 11.2 Optional accessories ...........................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Figure 25 Figure 26 865 Dosimat plus Front 865 Dosimat plus ..................................................................... 6 Rear 865 Dosimat plus ..................................................................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 Instrument description The 865 Dosimat plus is a universally utilizable dosing instrument. Methods can be generated and stored under a new name. When a USB memory stick is connected as an external storage medium, then the methods can be exported to that USB memory stick. This function makes it possible for you to copy methods quickly and easily from one instrument to another.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 Intended use 1.2 Intended use The 865 Dosimat plus is designed for use in laboratories and production plants. Its main area of application is the precise dosing of liquids. This includes simple dosing according to specific criteria, the preparing of solutions, the pipetting of liquid samples and manual titrations. The present instrument is suitable for dosing chemicals and flammable solvents.
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction Warning This symbol draws attention to a possible hazard due to electrical current. Warning This symbol draws attention to a possible hazard due to heat or hot instrument parts. Warning This symbol draws attention to a possible biological hazard. Caution This symbol draws attention to a possible damage of instruments or instrument parts. Note This symbol marks additional information and tips. 1.4 Safety instructions 1.4.
■■■■■■■■■■■■■■■■■■■■■■ 1.4 Safety instructions Warning Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched. There are no parts inside the housing which can be serviced or replaced by the user. Mains voltage Warning An incorrect mains voltage can damage the instrument. Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
■■■■■■■■■■■■■■■■■■■■■■ 1.4.4 1 Introduction Flammable solvents and chemicals Warning All relevant safety measures are to be observed when working with flammable solvents and chemicals. ■ ■ ■ ■ 1.5 Set up the instrument in a well-ventilated location. Keep all sources of flame far from the workplace. Clean up spilled fluids and solids immediately. Follow the safety instructions of the chemical manufacturer.
■■■■■■■■■■■■■■■■■■■■■■ 2.1 865 Dosimat plus 2 Overview of the instrument 2.1 865 Dosimat plus 1 2 3 4 5 Figure 1 Front 865 Dosimat plus 1 Piston rod Of the dosing drive. 2 Contact pins For the data chip. 3 Coupling For switching the flat stopcock.
■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 1 3 4 2 5 Figure 2 Rear 865 Dosimat plus 1 Type plate Contains specifications concerning mains voltage and serial number. 2 Mains connection socket 3 USB (OTG) connector For connecting printers, USB sticks, USB hubs, etc. 4 MSB connector Metrohm Serial Bus. 5 For connecting a stirrer or an 805 Dosimat. Mini DIN, 9-pin. Remote connector For connecting the Manual Dosing Controller or instruments with a remote interface. D-Sub, 9-pin.
■■■■■■■■■■■■■■■■■■■■■■ 2.2 Manual Dosing Controller 2.2 Manual Dosing Controller 1 2 3 4 Figure 3 Manual Dosing Controller 6.2107.100 1 Dosing key To start the dosing. 2 Status LED Indicates the readiness of the instrument. 3 Stop key/Fill key For stopping and filling the dosing cylinder. 4 Connection cable With 9-pin D-Sub plug.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 3 Installation 3.1 Setting up the instrument 3.1.1 Packaging The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument. 3.1.2 Checks Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note. 3.1.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 Connecting the Manual Dosing Controller Caution Make sure that the flat side of the plug matches the marking on the socket. 3.3 Connecting the Manual Dosing Controller Figure 5 Connecting the Manual Dosing Controller The Manual Dosing Controller is connected to the remote connector on the rear of the instrument. 3.4 Connecting a balance Balances are equipped with a serial RS-232 interface as a rule. To connect a balance, you require a 6.2148.030 RS-232/USB Box.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation The parameters for the RS-232 interface on the instrument must match those on the balance (see "Editing the COM1 settings", page 35). Additionally consult the user manual of the balance. 3.5 Connecting a keyboard, printer and other USB devices The 865 Dosimat plus has a USB (OTG) connector. Use the provided 6.2151.100 adapter USB MINI (OTG) - USB A for connecting USB devices as e.g. printers, keyboards or USB sticks, see the following figure. 6.2151.
3.5 Connecting a keyboard, printer and other USB devices ■■■■■■■■■■■■■■■■■■■■■■ Note Most of the USB devices need a so-called hub in order to work correctly. A USB hub is a distributor to which several USB devices can be connected. USB hubs are available in specialty stores in a number of different models. The USB (OTG) connector of the 865 Dosimat plus has no such hub. The 6.2147.000 numerical USB keypad has a USB hub and two USB connectors. The following devices can only be connected to a 6.2147.
■■■■■■■■■■■■■■■■■■■■■■ USB stick 3 Installation USB MINI (OTG)-USB 6.2151.100 Figure 9 Printer Keypad 6.2147.000 Connecting the 6.2147.000 USB keyboard with USB stick and printer USB MINI (OTG)-USB USB-Hub USB stick 6.2151.100 Figure 10 3.6 RS-232/USB Box 6.2148.030 Printer Connecting the USB hub with USB stick, printer and the 6.2148.030 RS-232/USB Box (for connecting balances).
■■■■■■■■■■■■■■■■■■■■■■ 3.7 Connecting the 805 Dosimat 3.7 Connecting the 805 Dosimat For tandem operation (XDOS mode), an 805 Dosimat can be connected to the MSB 1 socket on the rear of the 865 Dosimat plus. Figure 12 Connecting the 805 Dosimat Caution Make sure that the flat side of the plug matches the marking on the socket. If the MSB 1 connector is already occupied by a stirrer (e.g. 801 Stirrer), then the 805 Dosimat can be connected to the MSB socket of the stirrer.
■■■■■■■■■■■■■■■■■■■■■■ 3.8 3 Installation Attaching the exchange unit Figure 14 Attaching the exchange unit To attach the exchange unit, proceed as follows: 1 Slide the exchange unit from the front onto the 865 Dosimat plus and push all the way to the rear. It must snap in audibly.
■■■■■■■■■■■■■■■■■■■■■■ 4.1 Switching the instrument on and off 4 Operation 4.1 Switching the instrument on and off Switching on the instrument Proceed as follows: 1 ■ ■ ■ Press the red [FILL] key. The instrument is initialized and a system test performed. This process takes some time. If a buret unit has been attached, then a request appears to carry out the PREP function: All tubings and the cylinder are rinsed with the PREP (Preparing) function.
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation Switching off the instrument The instrument is switched off with the [FILL] key. The fact that the key needs to be pressed down for an extended time prevents accidental switch off. Proceed as follows: 1 ■ Keep the red [FILL] key pressed down for at least 3 s. A progress bar is displayed. If the key is released during this time, then the instrument will not be switched off. 4.2 Fundamentals of operation 4.2.
■■■■■■■■■■■■■■■■■■■■■■ 4.2 Fundamentals of operation 4.2.2 Structure of the dialog windows The current dialog title is displayed on the left-hand side of the title line. The current status of the system is displayed in the upper right-hand corner: ready The instrument is in normal status. busy A method has been started. hold A method has been paused. Some dialogs have a so-called function bar on the bottom line.
■■■■■■■■■■■■■■■■■■■■■■ 4.2.4 4 Operation Entering text and numbers In the editing dialog for text or numerical input you can select the individual characters with the arrow keys. Use [OK] to apply the character in the input field. The following functions are available: Editing function Description Accept The modification is applied and the editing dialog is exited. Cancel The editing dialog is exited without applying the modification. Clear The content of the input field is deleted completely.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 Methods 4.2.5 Selecting from a selection list In a selection list, select the individual entries with the arrow keys [⇧] and [⇩]. Accept the selection with [OK] or [BACK]. 4.3 Methods 4.3.1 Method templates The 865 Dosimat plus contains method templates which are already configured and which can be adjusted to individual requirements. The following method templates can be selected: DOS Manual interactive dosing with optional calculation of results. For manual titrations.
■■■■■■■■■■■■■■■■■■■■■■ 4.3.2 4 Operation Creating a new method Proceed as follows to create a new method: 1 Open the method table ■ In the main dialog, select Method and press [OK]. The method table opens: 2 Select the method template ■ In the function bar, select New and press [OK]. The list of method templates opens: 3 Load the method template ■ Select the desired template and press [OK]. The method template is now loaded and is displayed in the main dialog under Method.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 Methods The method table opens: 2 Modify/apply the method name ■ In the function bar, select Store and press [OK]. A method name will be suggested for new methods. If the method has already been saved once, then the method name will be displayed: Apply the name: ■ Press [BACK]. The method will be saved and the method table is displayed. Enter a new name: ■ ■ ■ Press [OK]. The text editor opens. Enter a method name (max. 12 characters) and apply with Accept or [BACK].
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation 2 Select a method ■ Select the desired method. 3 Load the method ■ In the function bar, select Load and press [OK]. The method is now loaded and is displayed in the main dialog under Method. 4.3.5 Exporting a method The methods can be exported on a connected USB stick. Note This function is possible only if a USB stick is connected as an external storage medium.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 Stirrer operation 3 Export the method ■ In the function bar, select Export and press [OK]. The method is being exported. The directory structure on the USB stick is listed in chapter 5.3, page 33. 4.4 Stirrer operation You can control a connected stirrer directly in the main dialog. Proceed as follows: 1 Set the stirring rate ■ ■ Use the arrow key [↓] to select the Stirrer entry. In the function bar, select the Stir- or Stir+ function.
■■■■■■■■■■■■■■■■■■■■■■ 4.5 4 Operation Entering sample data Menu ▶ Sample data In the dosing modes DOS and CNT, a sample size can be used for calculation. The sample size can be entered using the keypad (in the main menu under Sample data) or transferred directly from a balance by means of a cable connection. When using direct data transfer from a balance, care must be taken to ensure that the settings for data transfer on the 865 Dosimat plus and on the balance are in agreement (see page 35).
■■■■■■■■■■■■■■■■■■■■■■ 4.7 Printing a report manually 2 Start the preparing Caution Make sure that the buret tip is directed into a vessel that can accommodate the cylinder volume of your buret unit several times over. ■ Select Yes and confirm the message with [OK]. Preparing is carried out. 4.7 Printing a report manually Menu ▶ Print reports To print a report manually, proceed as follows: 1 Open the main menu ■ In the main dialog, select Menu and press [OK].
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation 3 Select a report ■ Select the desired report and press [OK]. The report is being printed out. The following reports can be printed out manually: 865 Dosimat plus Results Result report with determination properties, sample data, calculated results, etc. Parameters Report with all method parameters of the loaded method. System System report with system settings, solution list, external devices, etc.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 Basic settings 5 System settings 5.1 Basic settings Menu ▶ System ▶ Settings This chapter contains a description of general instrument settings. User name A user name can be entered here for the report. This parameter will only be printed if a user has been defined. Input Default value max. 12 characters empty Instrument name An instrument name can be entered here for the report. This parameter will only be printed if a designation has been defined. Input Default value max.
■■■■■■■■■■■■■■■■■■■■■■ 5 System settings Note A second language must be installed in advance in order to be able to select it here. The installation may only be carried out by competent personnel. Dialog type The user dialog can be limited for routine operations. One can operate normally with methods in the limited dialog. However, no settings can be made or methods deleted. The resetting of the dialog does not take effect until the main menu is exited.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 Basic settings Contrast The contrast of the display can be adjusted with the arrow keys [⇦] and [⇨]. ■ ■ [⇦]: the contrast will be decreased by one step each time the key is pressed. [⇨]: the contrast will be increased by one step each time the key is pressed. Range Default value 150 ... 240 212 Note Alternatively, the contrast can also be modified in the following manner: Keep the red [STOP] key pressed down.
■■■■■■■■■■■■■■■■■■■■■■ 5 System settings 5.2 Managing solutions 5.2.1 General Menu ▶ System ▶ Solutions Solutions can be used in intelligent buret units or in non-intelligent buret units. Intelligent buret units have a built-in data chip on which the data for the reagent is stored. This data is automatically read out during attachment and entered in the solution list. The name and the type are specified for each solution in the solution list.
■■■■■■■■■■■■■■■■■■■■■■ 5.2 Managing solutions Cylinder volume Cylinder volume of the buret unit in mL. The cylinder volume is automatically read out with intelligent buret units. Selection Default value 1 | 5 | 10 | 20 | 50 20 Concentration Concentration of the solution. Range Default value –999999999 ... 9999999999 1.000 Concentration unit Unit of the concentration.
■■■■■■■■■■■■■■■■■■■■■■ 5 System settings Selection Default value on | off off Time interval This parameter is visible only when Monitoring = on. You will be notified that this time interval (in days) has elapsed when starting a method. You can then select whether or not you would still like to start the method. Range Default value 5.3 1 ... 999 d 999 d File management Menu ▶ System ▶ File management Note This menu item is visible only when a USB stick has been connected as an external storage medium.
■■■■■■■■■■■■■■■■■■■■■■ 5.4 Configuring external devices Directory structure on the USB stick A directory with the instrument number is generated on the USB stick. The structure within the directory appears as follows: Figure 16 Directory structure on the USB stick Backup All of the files of the backup are stored in this directory. The directory will be created the first time a backup is created. Files Exported methods will be stored in this directory.
■■■■■■■■■■■■■■■■■■■■■■ 5 System settings The printers that have the designation ESC-POS are so-called POS printers (point-of-sale printers), i.e. they print on continuous paper. Selection Default value Citizen (ESC-POS) | Custom (ESC-POS) | Epson | Epson (ESC-POS) | HP DeskJet | HP LaserJet | Seiko (ESC-POS) HP DeskJet Keyboard layout A commercially available USB keyboard can be connected to make it easier to enter text and numbers. Specify the country-specific keyboard layout here.
■■■■■■■■■■■■■■■■■■■■■■ 5.4 Configuring external devices Baud rate Transfer rate in characters per second. Selection Default value 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 9600 Data bits Number of data bits. Selection Default value 7|8 8 Stop bits Number of stop bits. Selection Default value 1|2 1 Parity Type of parity testing. Selection Default value even | none | odd none Handshake Type of data transfer protocol.
■■■■■■■■■■■■■■■■■■■■■■ 5 System settings 5.5 Instrument diagnosis 5.5.1 Loading program versions and language files Menu ▶ System ▶ Diagnosis New program versions or language files can be loaded from a USB stick. The corresponding file must be saved on the USB stick in a directory with the instrument number (e.g. 848 or 863). You can distinguish between language files and program files by noting how the file name is constructed. Program files They are instrument-specific.
■■■■■■■■■■■■■■■■■■■■■■ 5.5 Instrument diagnosis 3 Open the file selection ■ Press [OK]. The selection list with the program and language files available on the USB stick is opened. 4 Select the file ■ ■ Select the required file with the arrow keys. Press [OK]. 5 Start the update ■ Press [START]. The update process is started, it runs automatically. At the end of the process, the instrument is automatically switched off and then back on again. No user intervention is required. 5.5.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes 6 Parameters and dosing modes 6.1 Manual dosing (DOS) 6.1.1 General description Usage The dosing mode DOS is particularly suitable for performing manual titrations with indicator. A result can be calculated automatically from the dosed volume and a result report can be printed out. Different calculation variables can be defined in advance as parameters. The parameters as a whole can be saved as a method and then later used where required.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 Manual dosing (DOS) The [FILL] key can be used at the end of a determination to initiate the filling of the dosing cylinder. This automatically initiates the calculation of the result and the printing out of a result report, if the appropriate settings have been made; see following chapters. The result of the calculation is shown under the volume display. The result and the dosing volume can be deleted by pressing the [FILL] key once more. 6.1.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes Dosing ramp The dosing ramp is a delay in the dosing rate at the time of the start of a dosing step. It is advantageous primarily for manual titrations with indicator, when small volume steps are to be dosed before the equivalence point. The selectable delay time determines after which amount of time (in seconds) the defined dosing rate is to be reached. 0 ... 10 s 0s Range Default value mL/min 90 60 30 2 Figure 17 6.1.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 Manual dosing (DOS) Note If a solution has been selected under Parameters, then the corresponding titer and the concentration will be read out from the solution data under System ▶ Solutions and used for calculations. If this is not the case, then calculations will be made with the Default value 1. Calculation status Activating the calculation. Selection Default value on | off off on The settings for the calculation can be entered as soon as the calculation has been activated.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes Input Default value 10 digits 1.0 Divisor Divisor for the calculation formula. Input Default value 10 digits 1.0 Blank The blank value is subtracted from the dosing volume. Range Default value 6.1.5 0.0000 ... 1000000 mL 0.0000 mL Reports Menu ▶ Parameters ▶ Reports The reports that will be printed out automatically in connection with a determination are defined under Reports.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Extended dosing (XDOS) Selection Default value off | on off on If the pulse control is activated, dosing is controlled via the remote interface. The [GO] key starts the pulse control and is deactivated afterwards. Autostart at power-on If this setting is activated, the pulse control is already activated when switching on the instrument or when loading the method. Manually starting the corresponding method is therefore not necessary.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes time of the output of the report. This requires that the name of the solution be specified under Menu ▶ Parameters. Performing a dosing Pressing the [GO] key causes the dosing to run automatically. Three dosing criteria are applied for this process. The specified dosing criteria are observed in all cases. The dosing criterion that is not defined in a specific case (volume, time or dosing rate) is calculated by the instrument and shown on the display.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Extended dosing (XDOS) Single Dosing with one instrument. Tandem Dosing with two Dosimats. Tandem operation An 805 Dosimat is connected to the MSB connector of the 865 Dosimat plus for tandem operation. Note Two exchange units with the same cylinder size must be used for tandem operation. The principle of tandem dosing: Figure 18 Tandem operation The two Dosimats operate alternately during dosing.
■■■■■■■■■■■■■■■■■■■■■■ 6.2.4 6 Parameters and dosing modes Dosing parameters Menu ▶ Parameters ▶ Dosing parameters Dosing criteria Selection of the preset dosing criteria. Selection Default value Rate/Time | Volume/Rate | Volume/Time Volume/Rate Rate/Time Dosing is carried out at the selected dosing rate for the amount of time specified. The time is also running while the cylinder is being filled. Volume/Rate The volume specified is dosed at the selected dosing rate.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Extended dosing (XDOS) Selection Default value off off off No safety limit. Auto fill Automatic filling of the cylinder after the dosing. Selection Default value 6.2.5 off | on on Reports Menu ▶ Parameters ▶ Reports The reports that will be printed out automatically in connection with a determination are defined under Reports. Results The result report contains the calculated result and additional specifications.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes 6.3 Creating solutions (CNT D) 6.3.1 General description Usage The dosing mode CNT D (Content dosing) is suitable for the preparation of standard and other solutions. The 865 Dosimat plus automatically determines the volume of the solvent to be dosed on the basis of the sample size of the starting substance (solid substance or stock solution) and the specified target concentration.
■■■■■■■■■■■■■■■■■■■■■■ 6.3 Creating solutions (CNT D) where: = dosed volume in ml = sample size in g = optional factor, e.g. for non-100 % samples = conversion factor, depending on unit = target concentration in the selected unit = molar mass in g/mol = density of the dosed solvent in g/ml Table 1 Variables and constant values in the CNT D mode Type of calculation Unit Substance concentration mol/L 103 . . 1 Substance concentration mmol/L 106 . . 1 Mass concentration g/L 103 .
■■■■■■■■■■■■■■■■■■■■■■ 6.3.2 6 Parameters and dosing modes Selecting a solution Menu ▶ Parameters Solution Selection of the solution from the solution list. We recommend always selecting the solution. Solutions are defined under System ▶ Solutions. For exchange units with integrated data chip, a check is made in the method run to verify whether the correct solution has been attached. For exchange units without integrated data chip, only the cylinder volume is checked.
■■■■■■■■■■■■■■■■■■■■■■ 6.3 Creating solutions (CNT D) Density Density of the solvent to be dosed. If selecting unit is: %, mmol/kg, mol/kg or ppm. Input 10 digits Range 0.000000001 ... 9999999999 g/mL Default value 1.0 g/mL Factor Multiplication factor for the calculation of the volume to be dosed. The factor can be used in general as a correction factor, e.g. for compensating for volume contraction or for specifying the starting concentration. Input Range Default value 6.3.4 10 digits 0.000000001 ...
■■■■■■■■■■■■■■■■■■■■■■ 6.3.5 6 Parameters and dosing modes Reports Menu ▶ Parameters ▶ Reports The reports that will be printed out automatically in connection with a determination are defined under Reports. Results The result report contains the calculated result and additional specifications. Selection Default value Label 1 | off | on off Label 1 A result report is printed out in a simple label format.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 Pipetting and diluting (LQT) 6.4 Pipetting and diluting (LQT) 6.4.1 General description Usage The dosing mode LQT (liquid transfer) is suitable for the pipetting and diluting of liquids. The set parameters can be saved as a method and then later used where required. Pipetting Pipetting is the transferring of a particular volume of a sample from one vessel to another.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes The run can be canceled with the [FILL] key. The dosing cylinder is filled automatically. Up to five steps are carried out. A text is displayed on the instrument as a demand note with each step. The readiness for carrying out the next step is indicated by the green light of the LED on the Manual Dosing Controller. Advance to the next step with the green [GO] key.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 Pipetting and diluting (LQT) 6.4.2 Selecting a solution Menu ▶ Parameters Solution Selection of the solution from the solution list. We recommend always selecting the solution. Solutions are defined under System ▶ Solutions. For exchange units with integrated data chip, a check is made in the method run to verify whether the correct solution has been attached. For exchange units without integrated data chip, only the cylinder volume is checked.
■■■■■■■■■■■■■■■■■■■■■■ 6 Parameters and dosing modes Diluting volume Volume of liquid for after rinsing or diluting the sample. Range Default value 6.4.4 0.0000 ... 9999.99 mL 0.0000 mL Dosing parameters Menu ▶ Parameters ▶ Dosing parameters Dosing rate The rate at which dosing takes place. The maximum dosing rate depends on the cylinder volume (see Chapter 8.1.1, page 61). Range Selection Default value 0.01 ... 150.00 mL/min max. max.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 Pipetting and diluting (LQT) PC/LIMS The PC/LIMS report is a machine-readable report with all of the data important for a determination. The PC/LIMS report can be saved as a TXT file on a USB storage medium or sent via an RS-232 interface to a LIMS. The output location is defined in the system settings (see "PC/LIMS report", page 34). The file name of the TXT file is constructed as follows: PC_LIMS_ReportID1-YYYYMMDD-hhmmss.txt.
■■■■■■■■■■■■■■■■■■■■■■ 7 Handling and maintenance 7 Handling and maintenance 7.1 General The 865 Dosimat plus requires appropriate care. Excess contamination of the instrument may result in malfunctions and a reduction in the service life of the sturdy mechanics and electronics of the instrument. Severe contamination can also have an influence on the measured results. Regular cleaning of exposed parts can prevent this to a large extent. Spilled chemicals and solvents must be removed immediately.
7.2 Quality Management and validation with Metrohm ■■■■■■■■■■■■■■■■■■■■■■ Note You can find information on the subjects of quality management, validation and maintenance as well as an overview of the documents currently available at www.metrohm.com/com/ under Support.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix 8 Appendix 8.1 Exchange unit 8.1.1 Maximum dosing and filling rate The maximum dosing rate and maximum filling rate for the exchange unit depend on the cylinder volume: Cylinder volume Maximum rate 1 mL 3.00 mL/min 5 mL 15.00 mL/min 10 mL 30.00 mL/min 20 mL 60.00 mL/min 50 mL 150.00 mL/min Independent of the cylinder volume, values ranging from 0.01 to 150.00 mL/min can always be entered.
■■■■■■■■■■■■■■■■■■■■■■ 8.2 Stirring rate 8.2 Stirring rate The stirring rate can be adjusted in steps of –15 to +15. The approximate rotational speed can be calculated with the following formula: Rotational speed/min (r/min) = 125 · stirring rate Example: Stirring rate set: 8 Rotational speed in rpm = 125 · 8 = 1000 r/min 2000 1500 1000 500 -15 Figure 19 8.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix Mettler AX For the Mettler AX balance, the fields that contain the sample identification or the method must be designated as follows: ■ ■ ■ 8.4 Designation for the field with the method name: METHOD Designation for the field with sample identification 1: ID1 Designation for the field with sample identification 2: ID2 USB devices Note USB peripheral devices that are to be connected directly must support the USB 1.0/1.1 (Full Speed) or USB 2.0 (High Speed) standard.
■■■■■■■■■■■■■■■■■■■■■■ 8.4 USB devices 8.4.2 Key assignment of a USB keyboard A commercially available USB keyboard can be connected to make it easier to enter text and numbers. The respective editing dialog must be opened for text and number inputs.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix Note Inexpensive printers are often designed solely for use with a PC and may not be equipped with one of the printer languages listed above. Such models are not suitable for this reason. 8.5 System initialization In very rare instances, it could happen that a faulty file system (e.g. because of a program crash) will lead to an impairment of program functioning. The internal file system needs to be initialized in such cases.
■■■■■■■■■■■■■■■■■■■■■■ 8.6 Remote interface 3 Confirm the initialization Note If the query is not confirmed within 8 s, then the procedure will be interrupted. Press [BACK] twice. ■ Initialization is started. The process takes approximately 80 s. The instrument will be automatically restarted after successful initialization. 8.6 Remote interface 8.6.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix Pin No. Assignment Function 6 0 volt (GND) 7 +5 volts 8 Input 0 Start/Go 9 Input 1 Stop/Fill Inputs +5 V approx. 5 kΩ Pull-up tp tp > 100 ms active = low, inactive = high Outputs Open Collector tp tp > 200 ms active = low, inactive = high IC = 20 mA, VCEO = 40 V +5 V: maximum load = 20 mA 8.6.
■■■■■■■■■■■■■■■■■■■■■■ 8.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix Dosing mode LQT Ready/EOD [GO] [GO] [GO] [GO] [GO] [GO] EOD Output 0 Output 1 Output 2 Pulse Output 3 Error Output 4 Start Input 0 Stop Figure 25 [Fill] Input 1 Remote status diagram LQT EOD = End of Determination 8.7 Pulse control The 865 Dosimat plus can be controlled by a control device via an electronic pulse control for simple dosing. Controlling is carried out via the 9pin remote interface.
■■■■■■■■■■■■■■■■■■■■■■ 8.8 Remote control via an RS-232 connection Pin No. Assigment Function 9 Input 1 Fill (low) Further details about the remote interface see Remote interface, page 66. Autostart pulse control In order to guarantee an unattended remote control of the 865 Dosimat plus, the parameter "Autostart at power-on" can be activated. The instrument thus starts the loaded method immediately after switching on and is ready to receive and carry out dosing pulses.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix 6.2134.040 Figure 26 Connecting the RS-232/USB Box to the PC Interface parameters We recommend the following parameters for the RS-232 interface COM2: ■ ■ ■ ■ ■ Baud rate: 19200 Data bits: 8 Stop bits: 1 Parity: none Handshake: software On the 865 Dosimat plus the interface parameters are set under Menu ▶ System ▶ External devices ▶ COM2 settings. Transfer protocol The data communication is synchronous. On each command there is an answer by the instrument.
■■■■■■■■■■■■■■■■■■■■■■ 8.8 Remote control via an RS-232 connection Command Function Comment $D Scan instrument status Answers: Ready;0 , Busy;0 , or Hold;0 (0 = no message) If a message on the instrument requires the interaction of the user, the answer of the status scan displays the corresponding message number. Example: Busy;010-119 = "Check buret unit" The message can be quit with [OK] or [Cancel], see below.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix Answer of the instrument Comment OK Command executed E1 Method not found E2 Invalid variable E3 Invalid command 865 Dosimat plus ■■■■■■■■ 73
■■■■■■■■■■■■■■■■■■■■■■ 9.1 Dosing drive 9 Technical specifications 9.1 Dosing drive 10000 steps per cylinder volume Resolution Exchange unit Cylinder volume ■ ■ ■ ■ ■ Fulfills ISO/DIN standard 8655-3 Accuracy 9.2 1 mL 5 mL 10 mL 20 mL 50 mL Interfaces USB (OTG) connector For connecting USB devices. MSB connector For connecting a stirrer or an 805 Dosimat. Remote connector For connecting instruments with a remote interface. 9.
■■■■■■■■■■■■■■■■■■■■■■ 9.4 9 Technical specifications Safety specification Design and testing According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1 Protection class I, degree of protection IP40 Safety instructions The documentation contains safety instructions which have to be followed by the user in order to ensure safe operation of the instrument. 9.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Reference conditions 9.7 Reference conditions Ambient temperature +25 °C ± 3 °C Rel. humidity ≤ 60 % Operating temperature status Instrument in operation at least 30 minutes Validity of the data After adjustment 9.
■■■■■■■■■■■■■■■■■■■■■■ 10 Conformity and warranty 10 Conformity and warranty 10.1 Declaration of Conformity This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company. Name of commodity 865 Dosimat plus Versatile dosing instrument with exchangeable buret units for the dispensing of liquids in laboratories.
■■■■■■■■■■■■■■■■■■■■■■ 10.2 Quality Management Principles Manufacturer Metrohm Ltd., CH-9101 Herisau/Switzerland Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality management system for development, production and sales of instruments and accessories for ion analysis. Herisau, 16 January 2009 10.2 D. Strohm A. Dellenbach Vice President, Head of R&D Head of Quality Management Quality Management Principles Metrohm Ltd.
■■■■■■■■■■■■■■■■■■■■■■ 10 Conformity and warranty Manufacture The measures put into practice in the production of our instruments guarantee a constant quality standard. Production planning and manufacturing procedures, maintenance of production means and testing of components, intermediate and finished products are prescribed. Customer support and service Customer support involves all phases of instrument acquisition and use by the customer, i.e.
11.1 Scope of delivery 865 Dosimat plus 2.865.0010 ■■■■■■■■■■■■■■■■■■■■■■ 11 Accessories Note Subject to change without notice. 11.1 Qty. 1 Scope of delivery 865 Dosimat plus 2.865.0010 Order no. Description 1.865.0010 865 Dosimat plus Univeral dispenser for use with exchange units. 1 6.2107.100 Manual dosing controller Manual dosing controller for dosing and manual titration with the Dosimat plus 1 6.2151.100 Adapter USB MINI (OTG) - USB A For connecting USB instruments.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 11 Accessories Order no. Description 6.2739.010 Wrench For exchange units. Length (mm): 1 6.3026.220 190 Exchange Unit 20 mL Exchange unit with integrated data chip with 20 mL glass cylinder and light protection. PCTFE/PTFE flat stopcock, FEP tubing connection, anti-diffusion buret tip and standard amber glass reagent bottle. Volume (mL): 20 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements.
11.2 Optional accessories 11.2 ■■■■■■■■■■■■■■■■■■■■■■ Optional accessories Order no. Description 2.141.0100 USB Thermal printer Neo's Compact printer with USB interface, paper width 60 mm (40 characters). Including 6.2151.120 USB cable. 2.801.0040 801 Stirrer with support Magnetic stirrer with base plate, support rod and electrode holder for use with Titrino plus, Titrandos, Sample Processors, 805 Dosimat and 780/781 pH Meters. With permanently attached cable for MSB (Metrohm Serial Bus). 2.805.
■■■■■■■■■■■■■■■■■■■■■■ Order no. Description 6.1562.060 Pipetting equipment for Manual Dosing Controller 11 Accessories Pipetting equipment for Manual Dosing Controller for 865 and 876 Dosimats. 6.1562.140 Pipetting tubing 10 mL for 6.1562.060 6.2134.120 Cable RS-232 to Mettler AX Connecting Metrohm RS-232 interface to Mettler AX 6.2141.
11.2 Optional accessories Order no. Description 6.2141.260 Remote cable Compact Sample Changer - Dosimat plus and Titrino plus ■■■■■■■■■■■■■■■■■■■■■■ Cable for remote connection between Compact Sample Changer and Dosimat plus and further devices (e.g. Titrino plus or Pump Station). 6.2147.000 Numerical USB keypad Numerical USB keypad for Titrino plus, Dosimat plus, 862 Compact Titrosampler and 869 Compact Sample Changer. Incl. USB Hub with two connectors. 6.2148.
■■■■■■■■■■■■■■■■■■■■■■ Order no. Description 6.3026.210 Exchange Unit 10 mL 11 Accessories Exchange unit with integrated data chip with 10 mL glass cylinder and light protection. PCTFE/PTFE flat stopcock, FEP tubing connection, anti-diffusion buret tip and standard amber glass reagent bottle. Volume (mL): 10 6.3026.250 Exchange Unit 50 mL Exchange unit with integrated data chip with 50 mL glass cylinder and light protection.
■■■■■■■■■■■■■■■■■■■■■■ Index Index Numbers/Symbols 805 Dosimat Connect ............................. 14 A Adapter Connect ............................. 11 Air gap ..................................... 56 Aspiration rate .......................... 57 B Backup ..................................... 33 Balance ........................ 39, 49, 62 Connect ............................. 10 Balance type ............................. 35 Blank value ............................... 43 Buret unit Prepare (PREP) ......
■■■■■■■■■■■■■■■■■■■■■■ Result ....................................... 42 Routine dialog .......................... 29 RS-232 connection ................... 70 Run .......................................... 55 S Safety air gap ........................... 56 Safety instructions ...................... 3 Sample data Enter ...................... 25, 39, 49 Sample size ........................ 25, 49 Send from a balance ........... 62 Sample transfer ........................ 56 Serial number .....................