Owner's manual
2.2 Functions
Dosing Interface USB Toolbox, DLL Reference 17
2.2.16 DosContinue
DosContinue(long IfNo, long MsbNo)
Continue a halted Dosino action.
Input parameters
IfNo long Dosing Interface [1… ]
MsbNo long Dosino at MSB[1…4]
Output parameters
eReturnstate return value
2.2.17 DU_Prep
DU_Prep(long IfNo, long MsbNo, long InPort, float InVolume,
float InRate, long OutPort, float OutVolume, float OutRate,
long SpecPort, float SpecVolume, float SpecRate)
Prepares a Dosing Unit for further use. A "Prep" cycle includes emptying the
dosing cylinder and rinsing and filling the tubings in one automated process.
Input parameters
IfNo long Dosing Interface [1… ]
MsbNo long Dosino at MSB[1…4]
InPort long Filling port [1…4]
InVolume
float Fill tube volume [0…20000 µL]
InRate float Filling rate [0.01…166 mL/min]
OutPort long Output port 1 [1…4]
OutVolume float Dosing volume 1 [0…20000 µL]
OutRate float Dosing rate 1 [0.01…166 mL/min]
SpecPort long Output Port 2 [1…4]
Specvolume float Dosing volume 2 [0…20000 µL]
SpecRate float Dosing rate 2 [0.01…166 mL/min]
Output parameters
eReturnstate return value
2.2.18 Empty
Empty(long IfNo, long MsbNo, long InPort, float InVolume, float
InRate, long OutPort, float OutVolume, float OutRate, long
SpecPort, float SpecVolume, float SpecRate)
Empties the Buret Unit. Dosing cylinder and tubings are emptied in one auto-
mated process.
Input parameters
IfNo long Dosing Interface [1… ]
MsbNo long Dosino at MSB[1…4]
InPort long Filling port [1…4]
InVolume
float Fill tube volume [0…20000 µL]