815 Robotic USB Sample Processor XL Manual 8.815.
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 815 Robotic USB Sample Processor XL Manual 8.815.
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above. Documentation in additional languages can be found on http://products.metrohm.com under Literature/Technical documentation.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1.1 1 The 815 Robotic USB Sample Processor XL in the Titrando system .................................................................... 1 1.2 Instrument description ......................................................... 2 1.2.1 Model versions ........................................................................ 3 1.2.2 Instrument components .......................................................... 4 1.2.3 Intended use ..
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 3.9 Guide chain for cables and tubing .................................... 29 3.10 Equipping the titration head ............................................. 31 3.11 Connecting the tower stirrer ............................................. 33 3.12 Connecting an external pump ........................................... 34 3.13 Connecting MSB devices ................................................... 35 3.13.1 Connecting dosing devices .....................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 7.4 Mains connection ............................................................... 57 7.5 Safety specifications ........................................................... 57 7.6 Electromagnetic compatibility (EMC) ................................ 57 7.7 Ambient temperature ......................................................... 58 7.8 Reference conditions .......................................................... 58 7.9 Dimensions ..................
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Figure 25 Figure 26 Figure 27 Figure 28 Figure 29 Figure 30 Figure 31 Figure 32 Figure 33 Figure 34 Figure 35 Figure 36 Figure 37 Figure 38 VI ■■■■■■■■ The Titrando system ................................................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 The 815 Robotic USB Sample Processor XL in the Titrando system The 815 Robotic USB Sample Processor XL is a component of the modular Titrando system. Operation is carried out by a Touch Control with touchsensitive screen ("Stand alone" titrator) or by a computer with a corresponding software. A Titrando system can contain numerous, various kinds of devices.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 Instrument description Up to three control devices (Titrandos, Dosing Interfaces, USB Sample Processors etc.) can be controlled via USB connection by the Touch Control. With the tiamo software the system can arbitrarily be extended with control devices. Updating the device software is described in the manual for the Touch Control or in the tiamo help, respectively. 1.2 Instrument description The 815 Robotic USB Sample Processor XL is a versatile instrument.
■■■■■■■■■■■■■■■■■■■■■■ 1.2.1 1 Introduction Model versions The 815 Robotic USB Sample Processor XL is available in the following versions with different components. 2.815.0010 1-Tower version ■ ■ ■ ■ ■ ■ ■ 1 membrane pumpe and 1 valve 1 connector for an external pump 1 stirrer connector (tower stirrer) 1 Swing Head connector 3 MSB connectors for dosing devices or stirrers 2 USB connectors 1 controller connection 2.815.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 Instrument description 2.815.0110 2-Tower version ■ ■ ■ ■ ■ ■ ■ 2 membrane pumps and 2 valves 2 connectors for external pumps 2 stirrer connectors (tower stirrer) 2 Swing Head connectors 3 MSB connectors for dosing devices or stirrers 2 USB connectors 1 controller connection 2.815.
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction ■ ■ ■ ■ 1.2.3 One stirrer connector per tower For connecting a rod stirrer (802 Stirrer) or a magnetic stirrer (741 Stirrer). Controller connection For connecting a PC or Touch Control. Two USB connectors For connecting a printer, barcode reader or other control devices (Titrando, Dosing Interface etc.). Three MSB connectors (Metrohm Serial Bus) For connecting dosing devices (Dosimat with exchange unit or Dosino with dosing unit), stirrers or Remote Boxes.
■■■■■■■■■■■■■■■■■■■■■■ 1.4 Safety instructions Warning This symbol draws attention to a possible life hazard or risk of injury. Warning This symbol draws attention to a possible hazard due to electrical current. Warning This symbol draws attention to a possible hazard due to heat or hot instrument parts. Warning This symbol draws attention to a possible biological hazard. Caution This symbol draws attention to a possible damage of instruments or instrument parts.
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction Warning Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched. There are no parts inside the housing which can be serviced or replaced by the user. Mains voltage Warning An incorrect mains voltage can damage the instrument. Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
■■■■■■■■■■■■■■■■■■■■■■ 1.4 Safety instructions 1.4.4 Personnel safety Warning Wear protective goggles and working clothes suitable for laboratory work while operating the 815 Robotic USB Sample Processor XL. It is also advisable to wear gloves when caustic liquids are used or in situations where glass vessels could break. Warning Always install the safety shield supplied with the equipment before using the instrument for the first time. Pre-installed safety shields are not allowed to be removed.
■■■■■■■■■■■■■■■■■■■■■■ 1.4.5 1 Introduction Flammable solvents and chemicals Warning All relevant safety measures are to be observed when working with flammable solvents and chemicals. ■ ■ ■ ■ 1.4.6 Set up the instrument in a well-ventilated location. Keep all sources of flame far from the workplace. Clean up spilled liquids and solids immediately. Follow the safety instructions of the chemical manufacturer.
2.1 Front and rear ■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 2.
■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 1 Safety shield (6.2751.100) other models, see chap. Accessories. 2 Sample rack (6.2041.800) other models, see chap. Accessories. 3 Turntable With guide bolts. 4 Stirrer rail For magnetic stirrer (741 Stirrer). 5 Guide chain For cables and tubings. 6 Lift With titration head holder. 7 786 Swing Head (2.786.0040) Drive for a robotic arm. 8 Robotic arm (6.1462.050) With aspiration and dosing tips.
■■■■■■■■■■■■■■■■■■■■■■ 2.2 Rear panel 1 Distributor For rinsing equipment. 2 Membrane pump Pump 1. 3 Pump connection Pump 2. For the external pump. 4 Pump valve 5 Swing Head connector Mini DIN socket (9-pin). 6 Stirrer connector DIN socket. For rod stirrer (802 Stirrer) or magnetic stirrer (741 Stirrer). 7 Rear panel with connectors 8 Warning symbol (see Chapter 1.4.4, page 8) T1 Tower 1 With a 2-tower model.
■■■■■■■■■■■■■■■■■■■■■■ 2.3 2 Overview of the instrument Sample racks A sample rack is a turntable that acts as a receptacle for sample vessels. Various types of sample racks are available for different numbers and types of sample vessels. The 815 Robotic USB Sample Processor XL requires sample racks with up to a maximum of 48 cm diameter. Figure 5 6.2041.
■■■■■■■■■■■■■■■■■■■■■■ 3.1 Setting up the instrument 3 Installation 3.1 Setting up the instrument 3.1.1 Packaging The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument. 3.1.2 Checks Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note. 3.1.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 3 Installation Connecting a computer The 815 Robotic USB Sample Processor XL requires a USB connection to a computer in order to be able to be controlled by a PC software. When a 6.2151.000 controller cable is used, the instrument can be connected directly, either to a USB socket on a computer, to a connected USB hub or to a different Metrohm control instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 Connecting a computer For Windows Vista: The instrument is recognized and the installation assistant for the driver is started automatically. Select the option "Find and install driver software". Agree to all subsequent requests. The installation assistant will be exited automatically. Note The plug on the instrument end of the 6.2151.000 controller cable is protected with an anti-pull device to prevent the cable from being pulled out accidentally.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 3.4 Mounting the Swing Head to the Sample Processor 3.4.1 Mounting the Swing Head Dismounting the titration head holder 1 2 1 1 Loosen the screws of the holder on both sides. 2 Loosen and remove the holder from the holder plate of the guide chain. Use the hexagon key provided. The screws will be needed again later.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Mounting the Swing Head to the Sample Processor ■ Plug the Mini DIN plug into the socket 'Swing Head'. Mounting the Swing Head 4 1 2 3 1 Screw the Swing Head tightly to the holder plate of the guide chain. 2 Clamp the Swing Head between the guide jaws. 3 Screw the Swing Head to the guide jaws with the screws previously loosened. 4 Place the connection cable into the guide chain (see Chapter 3.4.2, page 19).
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 1 Figure 9 Mounting the tower extension 2 Mount the Swing Head to the holder plate of the guide chain. 2 3 Figure 10 Mounting the Swing Head to the tower extension 3 Screw the Swing Head tightly to the upper end of the tower extension. 3.4.2 Guide chain for cables and tubing Tubings and cables can be placed into the guide chain.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Mounting the Swing Head to the Sample Processor You can open the individual chain links with a screwdriver as follows. 1 Open the guide chain ■ ■ ■ ■ Insert a screwdriver into the groove located on the side of a chain link. Loosen the clip with a forceful leverage movement. Pull the clip out of the chain by hand. Repeat the above actions for each chain link.
■■■■■■■■■■■■■■■■■■■■■■ 3.4.3 3 Installation Mounting the robotic arm reinforcement In cases in which it is necessary to take up or strip off a filter or a pipetting tip on the robotic arm, you must mount a 6.2058.090 robotic arm reinforcement. The reinforcement provides the robotic arm with stable resistance when picking up or stripping the tools used and prevents it from bending. The reinforcement has to be placed before mounting a robotic arm. Proceed as follows: 1 2 6.2058.
■■■■■■■■■■■■■■■■■■■■■■ 3.5 Configuring the robotic arm 3.5 Configuring the robotic arm A variety of differently constructed robotic arms is available for enabling a wide range of applications. They differ from one another in their geometric sizes, e.g. swing radius or maximum permissible swing angle. The configuration data must be entered in the Sample Processor or in the control software prior to the assembly of the robotic arm. The data required is engraved on the underside of the robotic arm.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation The following figure illustrates the most important configuration data that needs to be set in the control software to ensure correct usage of a robotic arm (left-swinging, here). 1 2 3 4 5 Figure 14 Configuration data of the robotic arms 1 Swing axis This runs through the middle of the Swing Head drive. 2 Swing radius This is determined by the length of the robotic arm. The radius runs from the axis of rotation to the midpoint of the tip of the robotic arm.
■■■■■■■■■■■■■■■■■■■■■■ 3.6 Mounting the robotic arm Swing direction Left-swinging (swing direction +) or right-swinging (swing direction –) model versions are available as different types of robotic arms. Left-swinging means swinging from the initial position (pointing towards the middle of the rack) outwards to the left. In the case of a Sample Processor with two towers, a right-swinging robotic arm must be mounted on Tower 1, a left-swinging robotic arm on Tower 2.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation After the initialization of the Sample Processor, the drive disc of the Swing Head is positioned as though the robotic arm were located in the outermost position. 1 2 3 Figure 16 Mounting the robotic arm Mount the robotic arm as follows: 1 Place the 6.1458.040 titration head insert in the opening of the robotic arm and screw tight with the supplied screws. 2 Hold the robotic arm in such a way that the opening faces to the right.
■■■■■■■■■■■■■■■■■■■■■■ 3.7 Robotic arms with beaker sensor 3.7 Robotic arms with beaker sensor For safety reasons, the presence of a beaker on the sample rack of a Sample Processor can be detected. Some robotic arm model versions are therefore equipped with a beaker sensor. 1 2 3 6.1462.150 Figure 17 Connecting a beaker sensor (for example 6.1462.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation No separate LIFT command is required in such cases. Note The work position of the lift must be configured in such a way that the robotic arm is in place on the sample vessel. The robotic arm must bend very slightly while doing so, so that the Piezo sensor will generate a signal. 3.8 Installing rinsing and aspiration equipment Various tubings are necessary for rinsing the electrode and the dosing tips as well as for aspirating the sample solution after the titration.
■■■■■■■■■■■■■■■■■■■■■■ 3.8 Installing rinsing and aspiration equipment 3 Mount the feed line for the rinsing liquid ■ Remove the union nut of the left-hand connector of the distributor and guide it over the end of a 6.1812.000 PTFE tubing. You may have to extend the tubing end in order to be able to better mount the tubing, see note below. Pull the end of the tubing over the connection nipple of the distributor and fasten in place with the union nut.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation Mounting the distributor 2 1 6.1808.170 Figure 19 Mounting the distributor Proceed as follows: 1 Remove a chain link ■ Remove the clip of the third chain link of the guide chain. Pry out the clip with a screwdriver on both sides of the chain link, as shown in the preceding illustration. 2 Insert the distributor ■ Apply strong pressure to insert the 6.1808.170 distributor (with the tubing connected) into the open chain link. 3 Fix the rinsing tubings ■ 3.
■■■■■■■■■■■■■■■■■■■■■■ 3.9 Guide chain for cables and tubing Figure 20 Guide chain - Opening chain links 2 Insert into the guide chain ■ Place the required tubings or cables into the guide chain. 3 Close the guide chain ■ Close the clip for each chain link again by hand and apply forceful pressure to snap them into place. Caution Take care to ensure when mounting tubing and cables that there is no traction on the drives while moving the lift or swiveling the robotic arm.
■■■■■■■■■■■■■■■■■■■■■■ 3.10 3 Installation Equipping the titration head Mounting the aspiration and rinsing tubings 6.1805.060 6.1805.510 3 1 6.1543.170 Figure 21 2 Installing the rinsing tubings and the aspiration tip Proceed as follows: 1 Connect the rinsing nozzles ■ Connect the three rinsing tubings that are already connected to the Tower 2 distributor to the rinsing nozzles already mounted on the titration head. 2 Insert the aspiration tip ■ Insert the 6.1543.
■■■■■■■■■■■■■■■■■■■■■■ 3.10 Equipping the titration head Inserting the stirrer and the electrode, connecting the dosing tubings 6.1805.120 4 1.802.0010 6.2104.030 1 6.1236.020 3 6.0229.100 2 6.1909.050 The equipment of the titration head is completed as follows: 1 Insert the rod stirrer ■ ■ Insert the rod stirrer (802 Stirrer) into the rear opening of the titration head (at the arrow). Insert the cable into the guide chain. 2 Mount the stirring propeller ■ Fasten the 6.1909.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 4 Connect the dosing tubings ■ 3.11 Connect two dosing tubings (e.g 6.1805.120) to the pre-mounted dosing tips on the titration head. Connecting the tower stirrer A DIN socket for connecting a rod stirrer (802 Stirrer) or a magnetic stirrer (741 Stirrer) is located on the rear side of the tower.
■■■■■■■■■■■■■■■■■■■■■■ 3.12 Connecting an external pump 3.12 Connecting an external pump If no integrated pump or a Sample Processor model version without pumps is used, up to two external pumps per tower can be connected. The 843 Pump Station (as model version with membrane pumps or with peristaltic pumps) has two pump drives and is connected to two sockets of the Sample Processor via the 6.2141.300 connecting cable (double cable with two plugs).
■■■■■■■■■■■■■■■■■■■■■■ 3.13 3 Installation Connecting MSB devices In order to connect MSB devices, e.g.stirrers or dosing devices, Metrohm instruments are equipped with up a maximum of four connectors at what is referred to as the Metrohm Serial Bus (MSB). Various kinds of peripheral devices can be connected in sequence (in series, as a "daisy chain") at a single MSB connector (8-pin Mini DIN socket) and controlled simultaneously by the respective control instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3.13 Connecting MSB devices 3.13.1 Connecting dosing devices Three dosing devices can be connected to the instrument. The types of dosing devices that are supported are: ■ ■ ■ ■ 800 Dosino 700 Dosino 805 Dosimat 685 Dosimat Warning If a Dosino is connected to the 815 Robotic USB Sample Processor XL then the connection cable must be equipped with a T.2400.102 ferrite core.
■■■■■■■■■■■■■■■■■■■■■■ 3.13.2 3 Installation Connecting a stirrer or titration stand You can use a magnetic stirrer 801 Stirrer or 803 Ti Stand (stirring "from below") or the 804 Ti Stand with a rod stirrer 802 Stirrer (stirring "from above"). Connect a stirrer or a titration stand as follows: 1 Connect a stirrer or titration stand ■ ■ ■ ■ Exit the control software.
■■■■■■■■■■■■■■■■■■■■■■ 3.13 Connecting MSB devices 3.13.3 Connecting a remote box Instruments that are controlled via remote lines and/or which send control signals via remote lines can be connected using the 6.2148.010 remote box. In addition to Metrohm, other instrument manufacturers also use similar connectors that make it possible to connect different instruments together.
■■■■■■■■■■■■■■■■■■■■■■ 3.14 3 Installation Connecting USB devices Two USB connectors (Type A sockets) are available for connecting devices with USB interfaces. The 815 Robotic USB Sample Processor XL functions then as a USB hub (distributor). If you wish to connect more than two USB devices, you can also use an additional commercially available USB hub. Note When a USB device is connected, the control instrument recognizes which device is connected.
■■■■■■■■■■■■■■■■■■■■■■ 3.15 Mounting the base plate Switch the barcode reader to programming mode and make the following settings: 1 ■ ■ ■ ■ ■ 3.15 Select the keyboard layout for the desired country (USA, Germany, France, Spain, Switzerland (German)). This setting must match the setting in the control software. Make sure that the Ctrl characters (ASCII 00 to 31) are allowed to be sent.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation Mounting the base plate 6 4 5 3 2 1 6.2001.070 Figure 31 Mounting the base plate Proceed as follows: 1 Insert the enclosed countersunk screw from below into the opening on the very back of the base plate. 2 Place the washer (flat side upwards) over the countersunk screw and tighten the support rod with the countersunk screw. The necessary hexagon key is enclosed with the 815 Robotic USB Sample Processor XL.
■■■■■■■■■■■■■■■■■■■■■■ 3.16 Mounting the drip pan 5 Loosen the screw for fastening the bracing on the rear panel of the tower (see figure) and fix the bracing with the enclosed hexagon screw. 6 Fix the bracing to the support rod with a hexagon key, see figure. 3.16 Mounting the drip pan Serious damage to the instrument or a danger to the user can occur if chemicals or liquid samples are spilled. The use of the drip pan (6.2711.060) is recommended in order to avoid such incidents.
■■■■■■■■■■■■■■■■■■■■■■ 3.17 3 Installation Attaching the sample rack 1 Figure 33 Attaching the rack Put the rack into place as follows: 1 Carefully center the rack on the turntable. The guide bolts on the turntable must engage with the openings in the bottom of the rack. Tip: hold the rack in such a way that the printed Metrohm logo is legible horizontally. 2 Screw the fixing screw in the handle tight by turning it clockwise.
■■■■■■■■■■■■■■■■■■■■■■ 3.18 Mounting the safety shield 3.18 Mounting the safety shield Warning It is imperative that the safety shield be installed before the first time the 815 Robotic USB Sample Processor XL is used. The device is not permitted to be operated without a safety shield. 1 3 Figure 34 2 Mount the safety shield Proceed as follows: 1 Loosen the knurled screws on both sides of the tower. 2 Move the safety shield into position, starting from the top.
■■■■■■■■■■■■■■■■■■■■■■ 4 Handling and maintenance 4 Handling and maintenance 4.1 General The 815 Robotic USB Sample Processor XL requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the service life of the sturdy mechanics and electronics of the instrument. Severe contamination can also have an influence on the measured results. Regular cleaning of exposed parts can prevent this to a large extent.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 Sample Processor 5 Troubleshooting 5.1 Sample Processor Problem Cause Remedy The instrument is not recognized by the control software. Sample Processor – No USB connection available. 1. Correctly plug in the USB connection cable on both ends. 2. Restart the control software or switch the Touch Control off and on again. Sample Processor – Power supply of the instrument is missing. 1. Plug in the mains cable on the instrument. 2.
■■■■■■■■■■■■■■■■■■■■■■ Problem 5 Troubleshooting Cause Remedy Swing Head – The Swing Head drive is defective. Contact the Metrohm Service. Problem Cause Remedy The pump is leaking. Sample Processor – A tubing connection is leaking. Check the tubing connections especially between the distributor an the pump and seal tightly. Canister – There is too much pressure on the pump valve. ■ 5.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 Beaker sensor 6 Appendix 6.1 Beaker sensor Every tower of a Sample Processor is equipped with a beaker sensor detecting the availability of a sample vessel in front of the tower. An infrared sensor identifies devices of various materials if they are located in a correct position in the front of the tower. In the rack configuration of the control device or the control software, the setting 'Beaker sensor' Tower must be selected.
■■■■■■■■■■■■■■■■■■■■■■ 6 Appendix Figure 36 Spray nozzles - Functioning The height of the nozzles can be adjusted in the titration head in order to reach an optimal rinsing effect. 6.3 Remote interface The 6.2148.010 remote box allows devices to be controlled which cannot be connected directly to the MSB interface of the Sample Processor. 1 2 3 Figure 37 Connectors of the remote box 1 Cable For connecting the Sample Processor. 3 Remote connector For connecting devices with a remote interface.
■■■■■■■■■■■■■■■■■■■■■■ 6.3 Remote interface 6.3.1 Pin assignment of the remote interface 13 1 1 13 25 14 14 25 Figure 38 Pin assignment of the remote socket and plug The above presentation of the pin assignment of a Metrohm remote interface applies not only for the remote box, but also for all Metrohm devices with 25-pin D-Sub remote connection. Inputs +5 V approx. 50 kΩ Pull-up tp tp >20 ms active = low, inactive = high The input lines can be scanned with the SCAN command.
■■■■■■■■■■■■■■■■■■■■■■ 6 Appendix Assigment Pin No. Assigment Pin No. Input 6 24 Output 6 1 Input 7 12 Output 7 2 0 volts / GND 14 Output 8 6 +5 volts 15 Output 9 7 0 volts / GND 25 Output 10 8 Output 11 13 Output 12 19 Output 13 20 6.4 Robotic arms 6.4.1 Robotic arms for titration Table 2 Configuration data of the titration robotic arms Type 6.1462.050 6.1462.060 6.1462.070 6.1462.260 6.1462.070 6.1462.060 6.1462.050 6.1462.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 Robotic arms 6.1462.060 Robotic arm with holder for a titration head, left-swinging The arm can be modified to create the desired titration robotic arm by means of the installation of a titration head 6.1458.xxx. Material: PP 6.1462.070 Robotic arm with holder for a titration head, right-swinging The arm can be modified to create the desired titration robotic arm by means of the installation of a titration head 6.1458.xxx. Material: PP 6.1462.
■■■■■■■■■■■■■■■■■■■■■■ 6 Appendix The arm can be equipped with two microelectrodes, one propeller stirrer and three spray nozzles. Two buret tips with anti-diffusion valve and one aspiration tip with connections for M6 tubing are already retracted into the arm. Material: PP Robotic arm with holder for a titration head and beaker sensor, left-swinging 6.1462.160 The arm can be modified to create the desired titration robotic arm by means of the installation of a titration head 6.1458.xxx.
■■■■■■■■■■■■■■■■■■■■■■ 6.4 Robotic arms Robotic arm for fully automated pipetting or dilution of liquid samples with Sample Processor Systems. Material: PP Robotic arm with Luer lock adapter, right-swinging 6.1462.090 For the connection of hollow needles with Luer lock connection. Suitable for the transfer of samples from sealed vials with septum seal. Material: PP Robotic arm with transfer head, bent, right-swinging 6.1462.240 The transfer head can, when equipped with 6.1808.
■■■■■■■■■■■■■■■■■■■■■■ 6.1462.250 6 Appendix Robotic arm as holder for a Polytron, left-swinging The robotic arm makes it possible to use the Polytron for sample preparation on multirow sample racks. It contains one retracted buret tip for adding solvents and three spray nozzles for cleaning. Material: PP 6.1462.080 Robotic arm DIS-COVER, left-swinging Robotic arm for placing and removing sample vessel covers (75 and 250 mL) covers 6.2037.050 und 6.2037.
■■■■■■■■■■■■■■■■■■■■■■ 7.1 Lift and turntable 7 Technical specifications 7.1 Lift and turntable Stroke path 235 mm Maximum lift load approx. 30 N / 3 kg Lift rate adjustable, 5…25 mm/s Shift rate adjustable, 3...20 angle degrees/s 7.2 Membrane pump(s) with valve > 450 mL/min Pressure head 2 m Capacity 7.3 Interfaces and connectors Controller connection USB Upstream Port (9-pin Mini DIN socket) for connecting a computer for controlling of the instrument.
■■■■■■■■■■■■■■■■■■■■■■ 7.4 7 Technical specifications Mains connection Voltage 100…240 V Frequency 50…60 Hz Power consumption 115 W Fuse 2.0 ATH 7.5 Safety specifications Design and testing According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1, EN/IEC 61010-2-081, protection class Ⅰ Safety instructions This document contains safety instructions which have to be followed by the user in order to ensure safe operation of the instrument. 7.
■■■■■■■■■■■■■■■■■■■■■■ 7.7 Ambient temperature 7.7 Ambient temperature Nominal function range 5…45 °C Humidity < 80 % Storage –20…60 °C Transport –40…60 °C 7.8 Reference conditions Ambient temperature 25 °C (±3 °C) Relative humidity ≤ 60 % 7.9 Dimensions Width 0.28 m Height 0.73 m Depth 0.53 m Weight (without accessories) 1.815.0010: 14.52 kg 1.815.0020: 15.42 kg 1.815.0030: 13.82 kg 1.815.0110: 18.87 kg 1.815.0120: 19.93 kg 1.815.0130: 16.
■■■■■■■■■■■■■■■■■■■■■■ 8 Conformity and warranty 8 Conformity and warranty 8.1 Declaration of Conformity This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
■■■■■■■■■■■■■■■■■■■■■■ 8.2 Warranty (guarantee) EN 61010-2-081 Manufacturer Particular requirements for automatic and semiautomatic laboratory equipment for analysis and other purposes Metrohm Ltd., CH-9101 Herisau/Switzerland Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality management system for development, production and sales of instruments and accessories for ion analysis. Herisau, 20 February, 2009 8.2 D. Strohm A.
■■■■■■■■■■■■■■■■■■■■■■ 8.3 8 Conformity and warranty Quality Management Principles Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
8.3 Quality Management Principles ■■■■■■■■■■■■■■■■■■■■■■ tional Qualification and Performance Verification of the system components or in carrying out the System Validation for the quantitative determination of a substance in a given matrix.
■■■■■■■■■■■■■■■■■■■■■■ 9 Accessories 9 Accessories Note Subject to change without notice. 9.1 Qty. 1 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0010 Order no. Description 1.815.0010 815 Robotic USB Sample Processor XL (1T/ 1P) Highly efficient Sample Processor with one work station and one internal membrane pump. 2 6.1236.020 Sleeve with SGJ 14/12 mm Sleeve with SGJ 14/12 mm and O-ring. Material: PP 5 6.1446.
9.1 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0010 Qty. 3 Order no. Description 6.1446.010 Stopper Used with Sample Changers Material: Height (mm): Outer diameter (mm): 1 6.1621.000 ■■■■■■■■■■■■■■■■■■■■■■ PVDF 19 9.3 Container 10 L As rinsing or waste container in automated systems. Material: PE Width (mm): 265 Height (mm): 400 Volume (mL): 10000 3 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.1805.110 FEP tubing / M6 / 80 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.1812.000 PTFE tubing 4/6 mm, 4m Material: Outer diameter (mm): Inner diameter (mm): 1 6.1828.000 6.2053.000 PTFE 6 4 PVDF connection nipple For 6.1621.000 container Material: 1 FEP 2 800 PVDF Cable clip Cable clip for fastening cables and tubes 1 6.2151.
■■■■■■■■■■■■■■■■■■■■■■ 9.1 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0010 Qty. 1 Order no. Description 6.2621.030 Hexagon key 4 mm Length (mm): 1 6.2621.070 5 mm. Length (mm): 73 Hexagon key 5 mm 80 1 6.2621.140 Hexagon key 2.5 mm 3 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 3 9 Accessories Order no. Description 6.2740.020 Spray nozzle For the fine-spraying of the rinsing solution Material: ETFE Outer diameter (mm): 10 Length (mm): 47 1 6.2751.100 Splash protection Splash protection for Sample Processors with 786 Swing Head 4 T.240.0102 Ferrite cores Anti-interference adapters 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.
■■■■■■■■■■■■■■■■■■■■■■ 9.2 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0020 9.2 Qty. 1 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0020 Order no. Description 1.815.0020 815 Robotic USB Sample Processor XL (1T/ 2P) Highly efficient Sample Processor with one work station and two internal membrane pumps. 2 6.1236.020 Sleeve with SGJ 14/12 mm Sleeve with SGJ 14/12 mm and O-ring. Material: PP 5 6.1446.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.1446.160 Thermometer stopper, B-14/15, oblique insertion To have the inserts of the 6.1458.010 titration head in a sloping position. Material: PTFE Height (mm): 21 Outer diameter (mm): 13 SGJ size: B-14/15 1 6.1543.170 Aspiration tip / M8 thread Aspiration tip to Sample Processors Material: PTFE Length (mm): 198 2 6.1621.000 Container 10 L As rinsing or waste container in automated systems.
9.2 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0020 Qty. 3 Order no. Description 6.1805.060 FEP tubing / M6 / 60 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.1805.110 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 9 Accessories Order no. Description 6.1812.000 PTFE tubing 4/6 mm, 4m Material: Outer diameter (mm): Inner diameter (mm): 1 6.1820.070 PTFE 6 4 Coupler 4/6 mm / M8 inside Extension for aspiration tubings in automated systems. Connects the M8 aspiration tubing directly with the 6.1812.000 PTFE tubing. Material: PTFE 2 6.1828.000 PVDF connection nipple For 6.1621.000 container Material: 1 6.2042.
9.2 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0020 Qty. 1 Order no. Description 6.2151.000 Cable USB A – mini-DIN 8-pin Controller cable Length (m): 1 6.2621.030 Length (mm): 1 6.2621.070 5 mm. Length (mm): 1 72 ■■■■■■■■ 6.2621.140 ■■■■■■■■■■■■■■■■■■■■■■ 1.8 Hexagon key 4 mm 73 Hexagon key 5 mm 80 Hexagon key 2.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 3 9 Accessories Order no. Description 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.5 3 6.2740.020 Spray nozzle For the fine-spraying of the rinsing solution Material: ETFE Outer diameter (mm): 10 Length (mm): 47 1 6.2751.100 Splash protection Splash protection for Sample Processors with 786 Swing Head 4 T.240.0102 Ferrite cores Anti-interference adapters 1 6.2122.
■■■■■■■■■■■■■■■■■■■■■■ 9.3 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0030 Qty. 1 9.3 Qty. 1 Order no. Description 8.815.8003EN 815 Robotic USB Sample Processor XL Manual Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0030 Order no. Description 1.815.0030 815 Robotic USB Sample Processor XL (1T/ 0P) Highly efficient Sample Processor with one work station and two pump connectors. 2 6.1236.020 Sleeve with SGJ 14/12 mm Sleeve with SGJ 14/12 mm and O-ring.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.2621.030 Hexagon key 4 mm Length (mm): 1 6.2621.070 73 Hexagon key 5 mm 5 mm. Length (mm): 80 1 6.2621.140 Hexagon key 2.5 mm 3 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.
■■■■■■■■■■■■■■■■■■■■■■ 9.4 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0110 Qty. 1 Order no. Description 6.2751.100 Splash protection Splash protection for Sample Processors with 786 Swing Head 4 T.240.0102 Ferrite cores Anti-interference adapters 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 9.4 Qty. 1 8.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 4 9 Accessories Order no. Description 6.1236.020 Sleeve with SGJ 14/12 mm Sleeve with SGJ 14/12 mm and O-ring. Material: PP 10 6.1446.000 SGJ stopper / B-14/(15) Material: Height (mm): SGJ size: 6 6.1446.010 PP 30.5 B-14/(15) Stopper Used with Sample Changers Material: Height (mm): Outer diameter (mm): 2 6.1621.000 PVDF 19 9.3 Container 10 L As rinsing or waste container in automated systems.
9.4 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0110 Qty. 6 Order no. Description 6.1805.060 FEP tubing / M6 / 60 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 2 6.1805.110 6.1812.000 6.1828.000 PTFE 6 4 PVDF connection nipple For 6.1621.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 9 Accessories Order no. Description 6.2053.000 Cable clip Cable clip for fastening cables and tubes 1 6.2151.000 Cable USB A – mini-DIN 8-pin Controller cable Length (m): 1 6.2621.030 1.8 Hexagon key 4 mm Length (mm): 1 6.2621.070 73 Hexagon key 5 mm 5 mm.
■■■■■■■■■■■■■■■■■■■■■■ 9.4 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0110 Qty. Order no. Description 1 6.2621.140 Hexagon key 2.5 mm 6 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.5 6 6.2740.020 Spray nozzle For the fine-spraying of the rinsing solution Material: ETFE Outer diameter (mm): 10 Length (mm): 47 2 6.2751.
■■■■■■■■■■■■■■■■■■■■■■ Qty. Order no. 9 Accessories Description Anti-interference adapters 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 9.5 Qty. 1 8.815.8003EN 815 Robotic USB Sample Processor XL Manual Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0120 Order no. Description 1.815.
9.5 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0120 Qty. 10 Order no. Description 6.1446.000 SGJ stopper / B-14/(15) Material: Height (mm): SGJ size: 6 6.1446.010 PP 30.5 B-14/(15) Stopper Used with Sample Changers Material: Height (mm): Outer diameter (mm): 2 6.1446.160 ■■■■■■■■■■■■■■■■■■■■■■ PVDF 19 9.3 Thermometer stopper, B-14/15, oblique insertion To have the inserts of the 6.1458.010 titration head in a sloping position.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 4 9 Accessories Order no. Description 6.1621.000 Container 10 L As rinsing or waste container in automated systems. Material: PE Width (mm): 265 Height (mm): 400 Volume (mL): 10000 6 6.1805.060 FEP tubing / M6 / 60 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 2 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ 9.5 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0120 Qty. 2 Order no. Description 6.1805.510 PTFE tubing / M8 / 60 cm With kink protection Material: Inner diameter (mm): Length (mm): 4 6.1812.000 PTFE tubing 4/6 mm, 4m Material: Outer diameter (mm): Inner diameter (mm): 2 6.1820.070 PTFE 3 600 PTFE 6 4 Coupler 4/6 mm / M8 inside Extension for aspiration tubings in automated systems. Connects the M8 aspiration tubing directly with the 6.1812.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 9 Accessories Order no. Description 6.2042.020 Buret tip clip (magnetic stirrer) Clip for proper alignment of buret tips, dosing or aspiration tubings using a magnetic stirrer in the titration beaker of a sample changer 2 6.2053.000 Cable clip Cable clip for fastening cables and tubes 1 6.2151.000 Cable USB A – mini-DIN 8-pin Controller cable Length (m): 1 6.2621.030 1.8 Hexagon key 4 mm Length (mm): 1 6.2621.070 73 Hexagon key 5 mm 5 mm.
■■■■■■■■■■■■■■■■■■■■■■ 9.5 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0120 Qty. Order no. Description 1 6.2621.140 Hexagon key 2.5 mm 6 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.5 6 6.2740.020 Spray nozzle For the fine-spraying of the rinsing solution Material: ETFE Outer diameter (mm): 10 Length (mm): 47 2 6.2751.
■■■■■■■■■■■■■■■■■■■■■■ Qty. Order no. 9 Accessories Description Anti-interference adapters 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 9.6 Qty. 1 8.815.8003EN 815 Robotic USB Sample Processor XL Manual Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0130 Order no. Description 1.815.
9.6 Scope of delivery 815 Robotic USB Sample Processor XL 2.815.0130 Qty. 2 Order no. Description 6.1805.110 FEP tubing / M6 / 80 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.2151.000 Controller cable Length (m): 1 6.2621.030 Length (mm): 1 6.2621.070 5 mm. Length (mm): 88 ■■■■■■■■ ■■■■■■■■■■■■■■■■■■■■■■ FEP 2 800 Cable USB A – mini-DIN 8-pin 1.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 9 Accessories Order no. Description 1 6.2621.140 Hexagon key 2.5 mm 6 6.2709.070 Guiding sleeve For fixation of buret and dosing tips in titration heads and electrode holders Material: ETFE Length (mm): 16.5 2 6.2751.100 Splash protection Splash protection for Sample Processors with 786 Swing Head 4 T.240.0102 Ferrite cores Anti-interference adapters 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 9.7 Qty. 1 Order no. Description 8.815.8003EN 815 Robotic USB Sample Processor XL Manual Optional accessories Order no. Description 2.772.0120 772 Pump Unit «aspirate» Peristaltic pump for direct connection to Sample Processors. The «Aspiration» version contains numerous accessories for aspirating off the titrated solution. Dimensions in mm (W/H/D): 100/210/240 1 2.786.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 2.786.0020 786 Swing Head with transfer head right Additional equipment for Robotic Sample Processors XL for transferring / pipetting from small sample vessels into larger titration vessels on the rack or into an external titration cell; this can be located to the right of the work station. 1 2.786.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 2.801.0010 801 Magnetic Stirrer Magnetic stirrer without stand for use with Titrino plus, Titrandos, Sample Processors, 805 Dosimats and 780/781 pH Meters. With permanently attached cable for MSB (Metrohm Serial Bus). 1 2.802.0020 802 Stirrer Rod stirrer for sample changer and Sample Processor. With 6.1909.020 Propeller stirrer 104 mm and fixed cable. 1 2.802.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 2.804.0040 804 Ti Stand with stand Titration stand and controller for 802 Rod Stirrer. 804 Ti Stand together with 802 Rod Stirrer provides an alternative to the magnetic stirrer. Ti Stand with base plate, support rod and electrode holder. 1 2.823.0020 823 Membrane Pump Unit «aspirate» Membrane pump for direct connection to Sample Processors. The «Aspiration» version contains numerous accessories for aspirating off solutions. 1 2.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 6.1458.030 KF titration head with 4 x M10 thread With 4 x M10 thread connectors. Material: PTFE Height (mm): 40 Outer diameter (mm): 68 1 6.1458.040 Titration Head, 3 x SGJ14 Titration head for 120 mL beakers. With 3 openings SGJ 14 and 4 openings 6.4 mm, 2 dosing tips included. Suitable for rinsing and aspiration equipment. Material: PTFE 1 6.1458.070 Titration head / 2 x SGJ14 Titration head for 75 mL beakers.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.1462.040 Robotic arm with transfer head for 786 Swing Head, right swinging Transfer robotic arm for liquid handling purposes Material: PP (black) 1 6.1462.050 Robotic arm with titration head for 786 Swing Head, left or right swinging Titration head for Swing Head Material: 1 6.1462.060 PP (black) Robotic arm with holder for titration head, left swinging Titration head holder to 786 Swing Head Material: PP (black) 1 6.1462.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 6.1462.150 Robotic arm with titration head and beaker sensor for 786 Swing Head, left or right swinging Titration head with beaker sensor for 786 Swing Head, 2 x M10, 1 x SGJ 14.5, 2 x buret tip Material: PP (black) 1 6.1462.160 Robotic arm with holder for titration head and sensor for 786 Swing Head, left swinging Titration head holder with beaker sensor for 786 Swing Head Material: PP (black) 1 6.1462.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.2001.070 Support stand for Sample Processors Support stand for Sample Processors, for external titrations and dilutions (IC) 1 6.2041.350 Sample rack 48 x 75 mL Sample rack for 48 x 75 mL samples, two-row Material: ABS Outer diameter (mm): 420 Hole diameter (mm): 35 1 6.2041.400 Sample rack 126 x 11 mL Sample rack for 126 x 11 mL samples, multi-row Material: PVC Outer diameter (mm): 420 Hole diameter (mm): 16 / 65 1 6.2041.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 6.2041.440 Sample rack 148 x 11 mL Sample rack for 148 x 11 mL samples with 3 rinsing beakers Material: PVC Outer diameter (mm): 420 Hole diameter (mm): 17 / 68 1 6.2041.450 Sample rack 56 x 11 mL + 56 x 50 mL Sample rack 56 x 11 mL + 56 x 50 mL Material: PVC Outer diameter (mm): 420 Hole diameter (mm): 29 / 17 1 6.2041.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.2041.820 Sample rack 28 x 250 mL Sample rack for 28 x 250 mL samples (6.1432.320), multi-row Material: PVC Outer diameter (mm): 480 Hole diameter (mm): 65 1 6.2041.830 Sample rack 28 x 200 mL Sample rack for 28 x 200 mL samples (6.1459.310), multi-row Material: PVC Outer diameter (mm): 480 Hole diameter (mm): 55 1 6.2041.840 Sample rack 59 x 120 mL Sample rack for 59x120 mL samples (6.1459.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 6.2058.020 Head extension 30 mm for Sample Processors Head extension 30 mm for 789, 815 Sample Processors. Material: Metal 1 6.2058.030 Head extension 10 mm for Sample Processors Head extension for use with the waterbath rack. Material: Metal 1 6.2141.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.2840.000 Water bath for Robotic Sample Processors and Robotic Titrosampler Water bath rack for Robotic Sample Processors and Titrosampler. Material: PVC 1 6.2841.000 External Polytron washing station for Robotic Sample Processors The external washing station can be mounted on the side of the Robotic Sample Processor tower and allows the external cleaning of the Polytron while other actions can be carried out on the sample rack.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 Optional accessories Qty. 1 Order no. Description 6.6056.211 tiamo™ 2.1 Light Program for controlling a titration system. Up to two instruments can be connected. Graphical method editor with numerous templates. Professional database with recalculation. Powerful report generator. No parallel titration, no data export. 1 License. Dialogue languages English, German, French, Italian, Chinese, Japanese, Portuguese, Russian, Slovak, Polish. 1 6.6056.212 tiamo™ 2.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 9 Accessories Order no. Description 6.6059.203 MagIC Net™ 2.0 Multi CD: 3 licences Professional PC program for controlling intelligent Professional IC systems, Compact IC instruments and their peripherals such as Professional Sample Processors, 800 Dosino, 771 Compact Interface, etc. The software permits control, data acquisition, evaluation and monitoring as well as report generation for ion chromatographic analyses.
■■■■■■■■■■■■■■■■■■■■■■ Index Index Numbers/Symbols 685 Dosimat ............................ 36 700 Dosino .............................. 36 741 Stirrer ................................ 33 800 Dosino .............................. 36 801 Stirrer ................................ 37 802 Stirrer ................................ 33 803 Ti Stand ............................. 37 804 Ti Stand ............................. 37 805 Dosimat ............................ 36 A Aspiration equipment ...............
■■■■■■■■■■■■■■■■■■■■■■ T Titration head ........................... 51 Equipping ........................... 31 Titration head holder ................ 17 Titration stand .......................... 37 Connecting ........................ 37 Tower extension ....................... 18 Tower stirrer ............................. 33 815 Robotic USB Sample Processor XL Index Transfer head ........................... 53 Type plate ................................ 12 V Validation .................................