3.8.1 Connecting dosing devices
Three dosing devices can be connected to the instrument.
The types of dosing devices that are supported are:
800 Dosino
700 Dosino
805 Dosimat
685 Dosimat
Warning
If a Dosino is connected to the 814 USB Sample Processor then the
connection cable must be equipped with a T.2400.102 ferrite core. The
ferrite core reduces any interference voltages that may occur and thus
ensures compliance with strict EMC standards pursuant to applicable
technical norms, see Chapter "Technical Data".
Proceed as follows:
1
Mounting ferrite core
Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.
2
Connect a dosing device
Exit the control software.
Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.
Start the control software.
USB 1
Contr.
MSB 2
MSB 3
T.2400.102
Figure 19 Connecting a dosing device