Owner's manual
4 Appendix
797 VA Computrace / Hardware-Manual 8.797.8001EN
86
Cable 6.2141.170 (option, for the 863 Compact Autosam-
pler)
6.2141.180 (option, for 838 Advanced Sample
Processor)
Sample changer 863 Compact Autosampler (2.863.0020, option)
or
838 Advanced Sample Processor (2.838.0310, option)
Rinsing equipment 843 Pump Station
or
731 Relay Box (2.731.0010, option) with two
772 Pump units (2 x 2.772.0110, option) or two 823
Membrane Pump Units (2 x 2.823.0010, option),
Adapter cable (6.2160.010) and Rinsing equipment
(6.5323.010, Option).
Mains connection
Voltage 100...240 V
Frequency 50...60 Hz
Power consumption 120 W
Fuse 2 × 1.6 ATH (to be replaced by Metrohm Service only
using the same type).
Additional electronic overload protection.
Safety specifications
Construction/testing
According to EN/IEC/UL 61010-1, CSA-C22.2 No.
61010-1
protection class 1
Safety directions The Instructions for Use include information and
warnings, which must be heeded by the user to assure
safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference Standards complied with:
- EN/IEC 61326-1
- EN 55022 / CISPR 22
- EN/IEC 61000-3-2
- EN/IEC 61000-3-3
Immunity to interference
Standards complied with:
- EN/IEC 61326-1
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6
- EN/IEC 61000-4-11
Ambient temperature
Nominal operating range 0…+45 °C