Owner's manual

2 Installation
797 VA Computrace / Hardware-Manual 8.797.8001EN
52
2 Set automation parameters
Select the Automation tab.
Select „813/863 Compact Autosampler“ for the field Sample processor in
the Sample handling-part of the
Automation tab.
Check the checkbox
Relay box.
Normally, the default settings can be used for samples with 10 mL volume:
3 Test automation parameters
Fill two sample vessels with water and place them in position 1 and 2 on
the sample rack of the 863 Compact Autosampler.
Click on
797 VA COMPUTRACE / Settings / General settings and select the Auto-
mation
tab.
Click on
, check the automation parameters and modify them if
needed.
2.9.4 Operation of the 863 Compact Autosampler
After installation of the instruments according to sections 2.9.1...2.9.3 sample series
using the 863 Compact Autosampler can be started. Proceed always in the following
sequence:
1 Switch on instruments
Switch on PC.
Switch on 797 VA Computrace Stand.
Switch on 863 Compact Autosampler and 843 Pump Station.