2.8 Connection of Dosing devices
797 VA Computrace / Hardware-Manual 8.797.8001EN
43
2 Insert 4-way microtip in 797 VA Computrace Stand (see section 2.2.4)
Remove stopper from nipple 27 and insert 4-way microtip 30 into nipple as
far as it will go (see Fig. 3 and Fig. 6).
Tighten nipple manually, so that the 4-way microtip can no longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip 26 in succession
from above through the opening
64.
3 Connect PTFE tubing to Exchange unit
Unscrew the attached 6.1805.020 FEP Tubing (L = 40 cm)
from connection 2 (connection for burette tip) of the flat
stopcock on the Exchange unit mounted on the Dosimat.
Screw connection nipple of the PTFE tubing of the
4-way microtip
30 onto connection 2 of the flat stopcock
on the Exchange unit mounted on the Dosimat.
2
4 Close unused PTFE tubing
Screw a 6.1808.000 Coupling (accessory of 797 VA Computrace Stand) on
each unused PTFE tubing of the 4-way microtip
30.
Screw a 6.1446.040 Dummy stopper (accessory of 797 VA Computrace
Stand) on each 6.1808.000 Coupling.
5 Initialize Dosimat(s)
Switch on the 797 VA Computrace Stand using the mains switch 14.
Switch on the PC and start the 797 VA Computrace program.
Open the
Dosinos tab (if the Dosino is connected to the 846 Dosing Interfa-
ce: the
Dosing Interface tab) of the window General settings and enter the
following configuration:
Note: The maximum allowed dose rate is 2 mL/min if the 4-way microtip is