Owner's manual

2 Installation
797 VA Computrace / Hardware-Manual 8.797.8001EN
42
Note: The maximum allowed dose rate is 2 mL/min if the 4-way microtip is
used to dose the solution in the measuring vessel.
Click <OK>.
Click on
or MAIN WINDOW / Utility / Dosino control to open the DOSINO
CONTROL
window.
The connected Dosino and the mounted Dosing unit are automatically iden-
tified.
Click the
button to empty and refill the Dosing unit installed on
the Dosino.
Check if there are air bubbles left in the glass cylinder of the Dosing unit. If
this is the case, repeat the flushing procedure by clicking the
button.
Close the
DOSINO CONTROL window.
Note: If you choose Port 3 for menu item
Prep / Empty via port, you must
install an FEP Tubing Connection 6.1805.XXX from Port 3 to a waste con-
tainer.
685 Dosimat and 805 Dosimat
1 Mount Exchange unit on Dosimat
Procedure, see Instructions for Use 685, section 2.1; and Instructions for
Use 805, section 3.1 respectively.