2.8 Connection of Dosing devices
797 VA Computrace / Hardware-Manual 8.797.8001EN
41
700 Dosino and 800 Dosino
1 Mount Dosing unit on Dosino
Procedure, see Instructions for Use 700, section 2.2, and Instructions for
Use 800, section 3.1 respectively.
2 Insert 4-way microtip in 797 VA Computrace Stand (see section 2.2.4)
Remove stopper from nipple 27 and insert 4-way microtip 30 into nipple as
far as it will go (see Fig. 3 and Fig. 6).
Tighten nipple manually, so that the 4-way microtip can no longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip 26 in succession
from above through the opening
64.
3 Connect PTFE-Tubing to Dosing unit
Screw connection nipple of the PTFE tubing of the
4-way microtip
30 on 1 (Port 1) of the Dosing unit
807.
4 Close unused PTFE tubing
Screw a 6.1808.000 Coupling (accessory of 797 VA Computrace Stand) on
each unused PTFE tubing of the 4-way microtip
30.
Screw a 6.1446.040 Dummy stopper (accessory of 797 VA Computrace
Stand) on each 6.1808.000 Coupling.
5 Initialize Dosino(s)
Switch on the 797 VA Computrace Stand using the mains switch 14.
Switch on the PC and start the 797 VA Computrace program.
Open the
Dosinos tab (if the Dosino is connected to the 846 Dosing Interfa-
ce: the
Dosing Interface tab) of the window General settings and enter the
following configuration: