Owner's manual

2.3 Multi-Mode Electrode (MME)
797 VA Computrace / Hardware-Manual 8.797.8001EN
33
Using a suitable coin, screw slotted screw 74 into the screw thread 80
until
the contact surface of the black O-ring at the Plexiglas wall (thin, black
stripe) is just visible below the metal thread.
Place Multi-Mode Electrode
21 with the opening 82 facing upwards in the
electrode holder
92 (see Fig. 11-1).
Insert ring
83 in opening 86.
Push sealing ring
84 onto locking ring 85 and insert both in the opening
82.
Screw retaining nut
86 by hand into screw thread 82 until a slight resis-
tance is felt.
9 Add mercury
Proceed as described in section 2.3.2.
10 Mount new capillary
Proceed as described in section 2.3.3.
11 Fill capillary
Proceed as described in section 2.3.4 or section 2.3.5.