5 Evacuating in vertical position
Place Multi-Mode Electrode 21 vertically in the electrode holder 92 (see
Fig. 11-1).
Evacuate for ca. 2 min in this position.
6 Evacuating in inclined position
Carefully tilt Multi-Mode Electrode 21 in the electrode holder 92 to an in-
clined position and continue evacuating (see
Fig. 11-2).
7 Release vacuum
As soon as mercury issues from the tip of the glass capillary 87 into the
filling tubing
93, carefully release the vacuum by opening the vacuum re-
lease tap.
The filling tubing
93 must not be disconnected from the glass capilla-
ry
87 when under vacuum, otherwise the mercury which has issued
from the capillary would be sprayed onto the tubing wall and can no
longer be disposed of in drop form!
Tap the glass capillary
87 gently by hand so that any mercury drops at its
tip are knocked into the filling tubing
93.
Disconnect filling tubing
94 with filling cone from glass capillary.
Place Multi-Mode Electrode
21 in a horizontal position in the electrode
holder
92 (see Fig. 11-3).
From now on, the MME must be left in this position until it is installed
in the stand!
8 Install Multi-Mode Electrode in 797 VA Computrace Stand
With measuring head arm 23 tilted back, push empty measuring vessel 7
into the holder
5 of the 797 VA Computrace Stand and then lower measur-
ing head arm
23.
Carefully insert Multi-Mode Electrode
21 in opening 58 of the measuring
head
24 (during insertion, the tip of the capillary 87 must not touch the
measuring head) and push in as far as it will go.