2.3.2 Filling the MME with mercury
When handling mercury, it is necessary to take special precautionary
measures. These are described in detail in section
3.4.
All actions involving the electrode and mercury vessels must be perfor-
med in or over the drip pan
91 supplied (see Fig. 9 - Fig. 11).
The Hg reservoir
81 of the Multi-Mode Electrode 21 is filled with mercury of the hig-
hest degree of purity (mass fraction w ≥ 0.99999) as follows:
1 Prepare Multi-Mode Electrode
• Unscrew locking ring 73 from slotted screw 74 (this gray PVC ring is nee-
ded only to remove the slotted screws
74 or 79, see section 2.3.7 and sec-
tion
2.3.9).
• Turn slotted screw
74 in or out of the screw thread 76 using a suitable
coin until the contact surface of the black O-ring at the Plexiglas wall (thin,
black stripe) is just visible below the metal thread
76.
• Remove the plastic cap used as a transport safeguard from the retaining
nut
86.
• Undo retaining nut
86 fully and remove from screw thread 82.
• Place Multi-Mode Electrode
21 with the capillary opening facing upwards
in the electrode holder
92 (see Fig. 9).
Fig. 9: Adding the mercury
21
Multi-Mode Electrode (6.1246.0020)
89
Syringe (6.2816.020)
90
Needle (6.2816.030)
91
Drip pan (6.2711.030)
92
Electrode holder (6.2615.030)
Hg
90
89
91
21
92