Instruction Manual

4 Exploratory mode
797 VA Computrace – Software
82
Charge (C)
Calculated charge transmitted during this peak (is
depending from sweep rate and current).
Edit baseline
Modify the peak evaluation of an already found peak. The results
are displayed in the PEAK SEARCH window. This window is opened
by clicking the number of a found peak in the PEAK SEARCH win-
dow with the right mouse button and selecting the menu point
Edit peak.
Begin
Manual setting of the start base point for baseline
calculation. The base point can be moved either
by manually changing the voltage value (time with
CPVS) in the first field or by clicking the
but-
tons of the second field indicating the current
value.
End
Manual setting of the end base point for baseline
calculation. The base point can be moved either
by manually changing the voltage value (time with
CPVS) in the first field or by clicking the
but-
tons of the second field indicating the current
value.
Type of baseline [ Linear, Polynomial, Exponential , Hori-
zontal ; Linear ]
Selection of the baseline type.
Scope [ Whole peak, Front end, Rear end ; Whole peak ]
Selection of the range for baseline evaluation.
This field can only be edited if the Linear baseline
type is selected.
Start peak evaluation with the current parameters
entered in the EDIT PEAK window.