Instruction Manual
4 Exploratory mode
797 VA Computrace – Software
80
Baseline Parameters for baseline evaluation.
Begin (V) (for CPVS: s) [ Start potential...End potential ; - ]
Manual setting of the start base point for baseline
evaluation. The values can be increased or de-
creased by clicking the buttons of the field or
by clicking the field and pressing the ↑ or ↓ key.
If the automatic peak evaluation is selected, n/a is
displayed and the field can not be edited.
End (V) (for CPVS: s) [ Start potential...End potential ; - ]
Manual setting of the end base point for baseline
evaluation. The values can be increased or de-
creased by clicking the buttons of the field or
by clicking the field and pressing the ↑ or ↓ key.
If the automatic peak evaluation is selected, n/a is
displayed and the field can not be edited.
Type [ Linear, Polynomial, Exponential , Horizontal ; Linear
]
Selection of the baseline type. Automatic peak
evaluation is only possible with baseline types lin-
ear and horizontal.
Scope [ Whole peak, Front end, Rear end ; Whole peak ]
Selection of the range for baseline evaluation.
This field can only be edited if the Linear baseline
type is selected.
Close the PEAK SEARCH window.
Clear all the peak evaluation results entered in the
peak table and the EXPLORATORY CURVES win-
dow.
Start peak evaluation with the current parameters
entered in the PEAK SEARCH window. The calcu-
lated baselines and peak maximum positions are
displayed in the
EXPLORATORY CURVES window.