Instruction Manual

2 Main window
797 VA Computrace – Software
32
3. After testing, the results are shown In the Dummy Cell Test_L
page of the
GLP WIZARD window.
4. Press <continue> one more time; the Dummy Cell Test_D page
of the GLP WIZARD window opens.
5. Proceed as described and press <continue> one more time; the
MONITOR window opens.
6. After testing, the results are shown In the Dummy Cell Test_D
page of the GLP WIZARD window.
7. Press <continue> one more time; the Summary of GLP valida-
tion
page of the GLP WIZARD window opens.
8. Press <Finish>.
Note: Metrohm recommends to check all 4 procedures on the
GLP
Wizard page of the GLP WIZARD window . The tests are progressed
in the order as listed.
GLP Electrode tests
1. If you check (only) the Electrode test checkbox on the GLP Wiz-
ard
page of the GLP WIZARD window and press <continue>,
you get to the Electrode Test page of the GLP WIZARD win-
dow.
2. Proceed as written and press <continue> one more time; the
Result part of the Electrode Test page of the GLP WIZARD win-
dow opens.
3. Press <continue> one more time; the Summary of GLP valida-
tion
page of the GLP WIZARD window opens.
4. Press <Finish>.
Note: Metrohm recommends to check all 4 procedures on the GLP
Wizard page of the GLP WIZARD window . The tests are progressed
in the order as listed.
GLP Validation of a chosen method
1. If you check (only) the Validation of a chosen method checkbox
on the GLP Wizard page of the GLP WIZARD window and press
<continue>, you get to the Validation of Standard Method
page of the GLP WIZARD window.
2. Choose the method Test Pb in standard solution.mth (or any
other predefined method from the Method folder) and press
<continue>.
3. Choose Number of measurements. The Analyte content is auto-
matically read from the method.